This study aimed to test the validity and reliability of an extended-field-of-view ultrasonography (EFOV) method for quadriceps muscle cross-sectional area (CSA) assessment. The CSA was measured at 10, 20, 30, 40 and 50% of the distance from the superior border of the patella to the medial aspect of anterior superior iliac spine by EFOV imaging and compared to the CSA measured by computed tomography (CT). Validity was tested by intra-class correlation (ICC) between the two methods and intra- and inter-experimenter and inter-day reliability were also examined. The ICC computed between the two techniques ranged between 0.951 and 0.998 (P < 0.000), but the 95% confidence intervals of the ICCs were in the acceptable range only for 30, 40 and 50% sections (0.842-0.999, 0.943-0.997 and 0.992-0.999, respectively). The standard error of the EFOV technique when compared to CT was 2.4, 4.3, 1.2, 1.2 and 0.6%, for 10, 20, 30, 40 and 50% sections, respectively. The coefficient of variation, showing intra- and inter-experimenter reliability, ranged from 0.6 to 2.7%. ICCs computed to assess the inter-day reliability were between 0.982 and 0.998 (95% confidence interval 0.892-1). When CSA was compared between sections statistically significant differences were found between them, regardless of the imaging technique used. Small standard errors of the measurement and high ICCs with the small confidence intervals suggest that, at proximal and mid-thigh sections, EFOV is a valid and reliable method to measure quadriceps muscle size.
本研究旨在测试一种扩展视野超声(EFOV)方法测量股四头肌横截面积(CSA)的有效性和可靠性。通过 EFOV 成像测量从髌骨上缘到髂前上棘内侧面的 10%、20%、30%、40%和 50%距离处的 CSA,并与 CT 测量的 CSA 进行比较。通过两种方法的组内相关系数(ICC)来测试有效性,还检查了实验内和实验间以及日内的可靠性。两种技术之间计算的 ICC 范围为 0.951 至 0.998(P<0.000),但 ICC 的 95%置信区间仅在 30%、40%和 50%节段处于可接受范围内(分别为 0.842-0.999、0.943-0.997 和 0.992-0.999)。与 CT 相比,EFOV 技术的标准误差分别为 10%、20%、30%、40%和 50%节段的 2.4%、4.3%、1.2%、1.2%和 0.6%。显示实验内和实验间可靠性的变异系数范围为 0.6%至 2.7%。用于评估日内可靠性的 ICC 范围为 0.982 至 0.998(95%置信区间为 0.892-1)。无论使用哪种成像技术,CSA 在节段之间的比较均存在统计学显著差异。测量的标准误差小且 ICC 高,置信区间小,这表明在大腿近端和中段,EFOV 是一种有效且可靠的测量股四头肌大小的方法。