School of Healthcare, University of Leeds, Leeds, UK.
Int J Audiol. 2010 Jun;49(6):463-6. doi: 10.3109/14992020903470791.
Audiologists typically verify hearing instrument fitting using real-ear measurements (REM). Recently the modified pressure with stored equalization method (MPSE) has been recommended for use when verifying open non-occluding hearing instruments. The MPSE method does not use a reference microphone to maintain loudspeaker output during real-ear measurements and is therefore susceptible to changes in the signal level at the client's ear which result from movement of the client's head and torso during the verification process. To determine the size of these errors, the real-ear unaided response (REUR) was measured prior to and following the fitting of a non-functioning hearing aid in the contralateral ear. Twenty young adults participated. Identical head positions for the two measurements should yield zero difference measures across all frequencies measured. Loudspeaker-to-client azimuths of 0 degrees and 45 degrees were investigated. Mean difference measures across the frequencies investigated were less than 1 dB for both azimuths with one standard deviation from these mean differences typically less than 1.5 dB. Results suggest that the MPSE method does not introduce clinically significant errors in real-ear measurements when verifying hearing instrument fitting in the population examined.
听力学家通常使用真耳测量(REM)来验证听力仪器的适配情况。最近,建议在验证开放式非闭塞听力仪器时使用存储均衡的改良压力(MPSE)方法。MPSE 方法在真耳测量期间不使用参考麦克风来维持扬声器输出,因此容易受到客户头部和躯干在验证过程中移动导致的客户耳朵处信号电平变化的影响。为了确定这些误差的大小,在对非工作助听器进行适配之前和之后测量了未受助的真耳响应(REUR)。二十名年轻人参与了这项研究。对于这两个测量,相同的头部位置应该在所有测量频率上产生零差异测量值。研究了扬声器到客户的 0 度和 45 度方位角。对于这两个方位角,所有研究频率的平均差异测量值均小于 1dB,这些平均差异的一个标准差通常小于 1.5dB。结果表明,在研究人群中验证听力仪器适配时,MPSE 方法不会在真耳测量中引入临床显著的误差。