Suppr超能文献

三维能量多普勒成像技术评估胎盘局部血流/灌注的可重复性研究。

Reproducibility of regional placental vascularity/perfusion measurement using 3D power Doppler.

机构信息

Department of Maternal-Fetal Medicine, Royal Prince Alfred Hospital, Camperdown, Australia.

出版信息

Ultrasound Obstet Gynecol. 2010 Aug;36(2):202-9. doi: 10.1002/uog.7608.

Abstract

OBJECTIVES

To assess reproducibility and regional variability of placental perfusion measurement using three-dimensional (3D) power Doppler VOCAL() (Virtual Organ Computer-aided AnaLysis).

METHODS

Twenty pregnant women at 26-34 weeks' gestation with normally grown, biophysically normal, singleton pregnancies with anterior placentae had placental power Doppler mapping data stored digitally from each of the four placental quadrants. Each was imaged by two investigators, with two datasets stored per investigator per quadrant. 5760 data values from the 320 datasets were evaluated by the same two investigators. Power Doppler imaging of the placental cord insertion was performed to generate a value for standardization as 'fractional moving blood volume' if appropriate. The vascularization index (VI), flow index (FI) and vascularization flow index (VFI) were calculated from spherical regions-of-interest to assess reproducibility within and between quadrants and between investigators for both acquisition and analysis.

RESULTS

We found extensive variability for all readings. For repeated measurements within the same dataset the intra-analyzer intraclass correlation coefficient (ICC) range was: 0.24-0.57 for VI, 0.33-0.78 for FI and 0.12-0.48 for VFI. The corresponding interanalyzer ICC range was: 0.38-0.92 for VI, 0.40-0.85 for FI and 0.10-0.92 for VFI. The intra-acquirer variability (paired t-test) mean differences range was: - 3.91 to 4.71 for VI, - 2.68 to 3.31 for FI and - 2.23 to 2.78 for VFI; the corresponding interacquirer variability (paired t-test) range was: - 1.92 to 5.18 for VI, - 3.06 to 2.04 for FI and - 1.69 to 2.60 for VFI. The regional variability range (coefficient of variation) was: 6.28-126.34% for VI, 2.26-49.01% for FI and 6.09-151.55% for VFI. For all analyzed data, FI showed least variability and cord values for VI were consistently 100% (mean VFI, 98.4 and 98.8 between observers).

CONCLUSIONS

There is insufficient evidence to support the meaning, reliability or reproducibility of VOCAL (VI, FI or VFI) as a tool to quantify placental perfusion, despite its use in multiple publications and journal submissions. There is poor reproducibility at the most fundamental level. Further investigation into the reproducibility of placental perfusion and quantification using VOCAL is required before development and application as a clinically useful tool.

摘要

目的

使用三维(3D)能量多普勒 VOCAL(虚拟器官计算机辅助分析)评估胎盘灌注测量的可重复性和区域变异性。

方法

20 名 26-34 周妊娠、正常生长、生物物理正常、前胎盘的单胎孕妇的胎盘彩色多普勒图谱数据以数字形式存储在每个胎盘的四个象限中。每个孕妇由两名研究人员进行成像,每个象限各存储两名研究人员的两个数据集。同两名研究人员评估 320 个数据集的 5760 个数据值。对胎盘脐带插入处进行彩色多普勒成像,以生成适当的“分数移动血容量”作为标准化值。从球形感兴趣区域计算血管化指数(VI)、血流指数(FI)和血管化血流指数(VFI),以评估象限内和象限间以及研究人员间采集和分析的重复性。

结果

我们发现所有读数都存在广泛的变异性。对于同一数据集内的重复测量,内部分析者组内相关系数(ICC)范围为:VI 为 0.24-0.57,FI 为 0.33-0.78,VFI 为 0.12-0.48。相应的外部分析者 ICC 范围为:VI 为 0.38-0.92,FI 为 0.40-0.85,VFI 为 0.10-0.92。内部获取者(配对 t 检验)差异均值范围为:VI 为-3.91 至 4.71,FI 为-2.68 至 3.31,VFI 为-2.23 至 2.78;相应的外部获取者(配对 t 检验)范围为:VI 为-1.92 至 5.18,FI 为-3.06 至 2.04,VFI 为-1.69 至 2.60。区域变异性范围(变异系数)为:VI 为 6.28-126.34%,FI 为 2.26-49.01%,VFI 为 6.09-151.55%。对于所有分析数据,FI 的变异性最小,VI 的脐带值始终为 100%(观察者之间的平均 VFI 为 98.4 和 98.8)。

结论

尽管 VOCAL(VI、FI 或 VFI)已在多篇出版物和期刊投稿中使用,但仍缺乏充分证据支持其作为量化胎盘灌注的工具的意义、可靠性或可重复性。在最基本的层面上,可重复性很差。在将其作为一种有临床应用价值的工具开发和应用之前,需要进一步研究使用 VOCAL 进行胎盘灌注和量化的可重复性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验