Division of Cardiology, University Health Network, Toronto, Ontario, Canada.
Heart Rhythm. 2010 Jun;7(6):852-5. doi: 10.1016/j.hrthm.2010.01.026. Epub 2010 Jan 22.
Two cases of "locked-in" sensitivity in the managed ventricular pacing mode are reported. Patient A was a 22-year-old man with a history of hypertrophic obstructive cardiomyopathy and a prophylactic dual-chamber automatic implantable cardioverter-defibrillator who manifested paradoxic T-wave oversensing on the ventricular lead at a sensitivity setting of 1.2 mV but not earlier when the sensitivity had been set to 0.9 mV. Patient B was a 74-year-old patient with ischemic cardiomyopathy and a dual-chamber defibrillator who presented with persistent paradoxic undersensing of the P wave on the atrial lead even at the maximal sensitivity of 0.15 mV, although P-wave sensing was normal earlier when sensitivity was set to 0.3 mV. The phenomenon and the mechanism for its behavior are described.
报告了两例在管理性心室起搏模式下的“锁定”感知现象。患者 A 为 22 岁男性,患有肥厚型梗阻性心肌病,预防性植入双腔自动植入式心脏复律除颤器。在感知灵敏度设置为 1.2 mV 时,心室导联出现了反常的 T 波过感知,但在灵敏度设置为 0.9 mV 时并未出现。患者 B 为 74 岁男性,患有缺血性心肌病和双腔除颤器。在感知灵敏度设置为 0.15 mV 时,心房导联始终存在持续性反常 P 波欠感知现象,尽管早些时候在感知灵敏度设置为 0.3 mV 时,P 波感知正常。描述了这种现象及其行为机制。