Technische Universität München, Garching, Germany.
Biomed Mater. 2010 Apr;5(2):25002. doi: 10.1088/1748-6041/5/2/025002. Epub 2010 Mar 8.
Injection-molded samples of thermoplastic polyetherurethane (TPU) were treated with low-temperature oxygen plasma for different processing times in order to enhance cellular attachment for a gastric implant. Its effects were investigated by contact angle measurement, surface topography, cytotoxicity and cell colonization tests. No significant changes were found in the surface roughness of plasma treatment with plasma treatment time of less than 5 min. Longer treatment showed significantly higher surface roughness. It seems that there was a link between the changes in contact angle and enhanced cell growth on the treated surface, although only for the range up to plasma treatment times of 3 min. Prolonged treatment times did not cause any major changes in the water contact angle, but strongly improved the number of growing cells on the surface. Plasma treatment for 3-7 min led to a twofold increase in the number of cells compared to untreated samples and did not significantly alter the WST-1 nor worsened the lactate dehydrogenase activity compared to the control. Thus, it appears that O(2) plasma treatment is a suitable surface modification method for a gastric implant made of TPU in order to improve surface cell attachment where 3-7 min is the recommended treatment time.
为了提高胃植入物的细胞附着能力,将热塑性聚醚聚氨酯(TPU)注塑样品用低温氧等离子体处理不同的处理时间。通过接触角测量、表面形貌、细胞毒性和细胞定植试验研究了其效果。在等离子体处理时间小于 5 分钟的情况下,表面粗糙度没有明显变化。更长的处理时间显示出明显更高的表面粗糙度。尽管仅在等离子体处理时间长达 3 分钟的范围内,接触角的变化与处理表面上增强的细胞生长之间似乎存在联系。处理时间延长不会导致水接触角发生重大变化,但会强烈增加表面上生长的细胞数量。与未处理的样品相比,处理 3-7 分钟会使细胞数量增加一倍,与对照相比,WST-1 或乳酸脱氢酶活性没有明显变化。因此,O(2)等离子体处理似乎是一种适合 TPU 胃植入物的表面改性方法,以提高表面细胞附着能力,其中 3-7 分钟是推荐的处理时间。