Department of Clinical and Biomedical Engineering, Rikshospitalet Oslo University Hospital, Sognsvannsveien 20, Oslo, Norway.
Ann Biomed Eng. 2010 Jul;38(7):2371-82. doi: 10.1007/s10439-010-9989-2. Epub 2010 Mar 9.
We present experimental findings regarding variability and stability of the electrical impedance properties of medical grade stainless steel needle electrodes in vitro. Monopolar impedance spectra (1 Hz to 1 MHz) were measured and scanning electron microscope images were obtained for five needle types with active electrode area from 0.28 to 0.7 mm(2). A saline tank (0.9% NaCl) was used as tissue model. Measurements were done before and after electrolytic treatment with anodic and cathodic DC currents of 1 muA. With active electrode areas below 1 mm(2), high influence from electrode polarization impedance (EPI) was expected at low frequencies (LF). For higher frequencies (HF) the EPI decreases and the impedance of the surrounding tissue is more pronounced. The hypothesis tested was that the EPI at LF would depend upon contact area, alloy composition, surface structure, and treatment of the active electrode, and at HF upon the electrode area geometry, and the specific resistivity of saline. Our results show large differences in electrical properties between needle types. After electrolytic treatment the EPI decreased. After 5-48 h of saline exposure the EPI increased, both for treated and untreated needles. Cathodic treatment gave lower impedance and drift than anodic or no treatment.
我们介绍了关于医用级不锈钢针电极体外电阻抗特性的变异性和稳定性的实验结果。测量了单极阻抗谱(1 Hz 至 1 MHz),并对五种针型进行了扫描电子显微镜图像分析,其有效电极面积从 0.28 到 0.7 mm²。盐水槽(0.9% NaCl)用作组织模型。在 1 μA 的阳极和阴极直流电流电解处理前后进行了测量。在有效电极面积低于 1 mm² 的情况下,低频(LF)下预计会受到电极极化阻抗(EPI)的高影响。在更高的频率(HF)下,EPI 降低,周围组织的阻抗更为明显。测试的假设是 LF 处的 EPI 将取决于接触面积、合金成分、表面结构以及有效电极的处理,而在 HF 处则取决于电极面积几何形状和盐水的比电阻率。我们的结果显示了不同针型之间的电性能存在很大差异。电解处理后,EPI 降低。盐水暴露 5-48 小时后,无论是处理过的还是未处理的针,EPI 都增加了。阴极处理比阳极处理或无处理具有更低的阻抗和漂移。