Validation of ultrafast computed tomographic left ventricular volume measurement.

作者信息

Pietras R J, Wolfkiel C J, Veselik K, Roig E, Chomka E V, Brundage B H

机构信息

Department of Medicine, University of Illinois, College of Medicine, Chicago.

出版信息

Invest Radiol. 1991 Jan;26(1):28-34. doi: 10.1097/00004424-199101000-00006.

Abstract

Left ventricular volume has been measured with ultrafast computed tomography. However, the accuracy with which this can be done is unknown. We therefore imaged with ultrafast computed tomography 11 rectangular phantoms, 20 to 225 ml, and 17 left ventricular casts, 15 to 112 ml. Two observers planimetered serial tomographic images and computed volume from sequential tomograms. There was no significant inter- or intraobserver difference in measurement of phantoms. Deviation of ultrafast computed tomographic volume from true phantom volume was -0.1 +/- 3.5% SD, range 9.0 to -7.6%. Correlation of true phantom volume with ultrafast computed tomographic volume was 0.99, SEE = 1.9 ml. No significant difference was observed between merged and single ultrafast computed tomographic scanning sequences. Left ventricular cast volume determined by ultrafast computed tomography deviated from true volume by 6% +/- 20%, range 54% to -45%. Correlation of true volume with ultrafast computed tomographic volume was 0.99, SEE = 5.1 ml. There was no interobserver significant difference in measurement of left ventricular cast volume. Correlation between ultrafast computed tomographic volume and cineradiographic volume of the same left ventricular casts was 0.99, SEE = 4.4 ml. Thus, phantom volumes can be measured accurately without significant intra- or interobserver variation. Merged scanning sequences did not influence volume determination. Left ventricular cast volume determination was comparable to that obtained with cineradiography.

摘要

已使用超快速计算机断层扫描测量左心室容积。然而,其测量的准确性尚不清楚。因此,我们使用超快速计算机断层扫描对11个容积为20至225毫升的矩形体模和17个容积为15至112毫升的左心室铸型进行成像。两名观察者对断层扫描图像进行了面积测量,并根据连续断层图像计算容积。在体模测量中,观察者间和观察者内均无显著差异。超快速计算机断层扫描容积与体模真实容积的偏差为-0.1±3.5%标准差,范围为9.0%至-7.6%。体模真实容积与超快速计算机断层扫描容积的相关性为0.99,标准误为1.9毫升。在合并和单次超快速计算机断层扫描序列之间未观察到显著差异。通过超快速计算机断层扫描确定的左心室铸型容积与真实容积的偏差为6%±20%,范围为54%至-45%。真实容积与超快速计算机断层扫描容积的相关性为0.99,标准误为5.1毫升。在左心室铸型容积测量中,观察者间无显著差异。同一左心室铸型的超快速计算机断层扫描容积与电影血管造影容积之间的相关性为0.99,标准误为4.4毫升。因此,可以准确测量体模容积,且观察者内和观察者间均无显著差异。合并扫描序列不影响容积测定。左心室铸型容积测定与电影血管造影获得的结果相当。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验