Precision of measurements of right and left ventricular volume by cine computed tomography.

作者信息

Reiter S J, Rumberger J A, Feiring A J, Stanford W, Marcus M L

出版信息

Circulation. 1986 Oct;74(4):890-900. doi: 10.1161/01.cir.74.4.890.

Abstract

Precise determination of left and right ventricular stroke volumes is limited with conventional imaging techniques. We determined whether right and left ventricular stroke volumes could be precisely measured with cine computed tomography (CT). Cine CT enables simultaneous imaging of the right and left ventricles at an 8 mm slice thickness with a maximal scanning rate of 17 frames/sec (50 msec acquisition intervals). In eight dogs, true right ventricular and left ventricular stroke volumes were determined by dividing thermodilution cardiac output by heart rate and/or with the use of an aortic electromagnetic flow probe implanted over a long term. After at least 5 sec of suspended respiration, cine CT images were acquired during central venous injection of a nonionic contrast agent. Multiple perturbations in stroke volume were induced in each dog by the administration of dobutamine, sodium pentobarbital, or sodium nitroprusside or by coronary artery occlusion. Right and left ventricular stroke volumes were obtained by Simpson's reconstruction of end-diastolic and end-systolic short-axis tomograms from apex to base. The cine CT left ventricular stroke volume (range 11 to 45 ml) correlated highly with the true left ventricular stroke volume (r = .99, slope = 1.01, y intercept = -0.2 ml, SEE = 1.5 ml, n = 25). The cine CT right ventricular stroke volume (range 11 to 34 ml) also correlated highly with the true right ventricular stroke volume (r = .98, slope = 0.9, y intercept = 2.2 ml, SEE = 1.7 ml, n = 15). In 12 studies, the mean difference between nearly simultaneous right and left ventricular stroke volumes by cine CT was 1.1 ml (range 0.1 to 3.2 ml). Calculation of right and left ventricular stroke volumes from data from cine CT were highly reproducible. Intraobserver variability in measurements of right ventricular stroke volume (r = 1.0, slope = 0.99, y intercept = 0.19 ml) and left ventricular stroke volume (r = 1.0, slope = 1.02, y intercept = -0.21 ml) was minimal. Interobserver variability in measurements of right ventricular stroke volume (r = .98, slope = 0.90, y intercept = 1.66 ml) and left ventricular stroke volume (r = .99, slope = 0.97, y intercept = -0.02 ml) was likewise minimal. Thus, precise and highly reproducible measurements of right and left ventricular stroke volumes can be obtained with cine CT.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验