Eur J Appl Physiol. 2010 Jul;109(4):779-86. doi: 10.1007/s00421-010-1415-9. Epub 2010 Mar 12.
The first aim of this study was to compare an ambulatory five-lead ECG system with the commercially available breast belt measuring devices; Polar S810i and Suunto t6, in terms of R-R interval measures and heart rate variability (HRV) indices. The second aim was to compare different HRV spectral analysis methods. Nineteen young males (aged between 22 and 31 years, median 24 years) underwent simultaneous R-R interval recordings with the three instruments during supine and sitting rest, moderate dynamic, and moderate to vigorous static exercise of the upper and lower limb. For each subject, 17 R-R interval series of 3-min length were extracted from the whole recordings and then analyzed in frequency domain using (1) a fast Fourier transform (FFT), (2) an autoregressive model (AR), (3) a Welch periodogram (WP) and (4) a continuous wavelet transform (CWT). Intra-class correlation coefficients (ICC) and Bland-Altman limits of agreement (LoA) method served as criteria for measurement agreement. Regarding the R-R interval recordings, ICC (lower ICC 95% confidence interval >0.99) as well as LoA (maximum LoA: -15.1 to 14.3 ms for ECG vs. Polar) showed an excellent agreement between all devices. Therefore, the three instruments may be used interchangeably in recording and interpolation of R-R intervals. ICCs for HRV frequency parameters were also high, but in most cases LoA analysis revealed unacceptable discrepancies between the instruments. The agreement among the different frequency transform methods can be taken for granted when analyzing the normalized power in low and high frequency ranges; however, not when analyzing the absolute values.
本研究的主要目的是比较动态五导联心电图系统与市售的胸带测量设备(Polar S810i 和 Suunto t6)在 RR 间期测量和心率变异性(HRV)指标方面的差异。次要目的是比较不同的 HRV 频谱分析方法。19 名年轻男性(年龄 22-31 岁,中位数 24 岁)在仰卧位和坐位休息、中强度动态运动以及中等到剧烈的上肢和下肢静态运动时,同时使用三种仪器进行 RR 间期记录。对于每个受试者,从整个记录中提取 17 个 3 分钟长度的 RR 间期序列,并使用(1)快速傅里叶变换(FFT)、(2)自回归模型(AR)、(3)Welch 周期图(WP)和(4)连续小波变换(CWT)在频域中进行分析。组内相关系数(ICC)和 Bland-Altman 一致性界限(LoA)方法作为测量一致性的标准。对于 RR 间期记录,ICC(较低的 ICC 95%置信区间>0.99)和 LoA(最大 LoA:ECG 与 Polar 相比为-15.1 至 14.3 ms)显示所有设备之间具有极好的一致性。因此,这三种仪器可在 RR 间期的记录和插值中互换使用。HRV 频域参数的 ICC 也很高,但在大多数情况下,LoA 分析显示仪器之间存在不可接受的差异。当分析归一化低频和高频范围的功率时,可以认为不同的频率变换方法之间具有一致性;但是,当分析绝对值时则不然。