de Yang L, Bairey C N, Berman D S, Nichols K J, Odom-Maryon T, Rozanski A
Department of Medicine, Cedars-Sinai Medical Center, Los Angeles, CA 90048.
J Nucl Med. 1991 May;32(5):796-802.
The accuracy and reproducibility of a new ambulatory radionuclide detector system (the VEST) for ejection fraction measurement has not been fully validated. Thirty-six subjects, (19 volunteers and 17 patients) underwent repetitive bicycle exercise using sequences of both VEST monitoring and gamma camera imaging. A high intraclass correlation was noted for both absolute ejection fraction [0.84 (0.56, 0.95)] and delta ejection fraction [0.87 (0.63, 0.96)] during repeat VEST monitoring. The intraclass correlation for ejection fraction was comparable for data averaged over 30 sec versus 2 min. These correlations compared favorably to those obtained for assessment of absolute and delta ejection fraction as derived by gamma camera determination by the same computer operator (intraobserver variability), two different computer operators (interobserver variability), and during repeat exercise using gamma camera imaging. In concordance, correlations between VEST and gamma camera measurements were relatively high for both absolute ejection fraction (0.78 [0.61, 0.88]) and delta ejection fraction (0.63 [0.39, 0.79]). Thus, the VEST represents a reproducible means of measuring ejection fraction change during dynamic physical activity. Its accuracy in ejection fraction measurements is similar to gamma camera imaging during exercise testing.
一种用于测量射血分数的新型动态放射性核素检测系统(VEST)的准确性和可重复性尚未得到充分验证。36名受试者(19名志愿者和17名患者)通过VEST监测和γ相机成像序列进行了重复性自行车运动。在重复VEST监测期间,绝对射血分数[0.84(0.56,0.95)]和射血分数变化值[0.87(0.63,0.96)]均呈现出较高的组内相关性。射血分数的组内相关性在30秒与2分钟的平均数据中相当。与由同一名计算机操作员通过γ相机测定得出的绝对射血分数和射血分数变化值评估(观察者内变异性)、两名不同计算机操作员(观察者间变异性)以及在使用γ相机成像的重复运动期间所获得的相关性相比,这些相关性表现良好。一致地,VEST与γ相机测量之间的相关性对于绝对射血分数(0.78 [0.61,0.88])和射血分数变化值(0.63 [0.39,0.79])均相对较高。因此,VEST是一种在动态体力活动期间测量射血分数变化的可重复方法。其在射血分数测量方面的准确性与运动测试期间的γ相机成像相似。