使用PTW Linacheck监测直线加速器输出稳定性。

Monitoring linear accelerator output constancy using the PTW Linacheck.

作者信息

McDermott Garry M, Buckle Andrew H

机构信息

Radiotherapy Physics Section, Department of Medical Physics and Clinical Engineering, Churchill Hospital, Oxford Radcliffe NHS Hospitals Trust, Headington, Oxford, UK.

出版信息

Med Dosim. 2011 Spring;36(1):71-4. doi: 10.1016/j.meddos.2009.12.001. Epub 2010 Mar 26.

Abstract

The PTW-Linacheck was assessed for its ability to monitor linear accelerator radiation output constancy. The key issues that were considered were the setup for daily output measurements, e.g., requirements for build-up and backscatter material, and the reproducibility and linearity of the device with linear accelerator output. An appropriate measurement setup includes a 10 × 10 cm field at 100 cm FSD, 5 cm backscatter, and no added build-up for 4 MeV electron beams, 1 cm added build-up for 6-16 MeV electron beams and 5 cm added build-up for 6-15 MV photon beams. Using this measurement setup, the dose linearity and short-term reproducibility were acceptable; however, the Linacheck should be recalibrated on a monthly basis to ensure acceptable long-term reproducibility.

摘要

对PTW-Linacheck监测直线加速器辐射输出稳定性的能力进行了评估。所考虑的关键问题包括每日输出测量的设置,例如对建成和散射材料的要求,以及该设备与直线加速器输出的再现性和线性。合适的测量设置包括在100 cm源轴距下的10×10 cm射野、5 cm散射、4 MeV电子束不添加建成材料、6 - 16 MeV电子束添加1 cm建成材料以及6 - 15 MV光子束添加5 cm建成材料。使用该测量设置,剂量线性和短期再现性是可接受的;然而,Linacheck应每月重新校准以确保可接受的长期再现性。

相似文献

1
Monitoring linear accelerator output constancy using the PTW Linacheck.
Med Dosim. 2011 Spring;36(1):71-4. doi: 10.1016/j.meddos.2009.12.001. Epub 2010 Mar 26.
2
Characterization of megavoltage electron beams delivered through a photon multi-leaf collimator (pMLC).
Phys Med Biol. 2006 Apr 21;51(8):2113-29. doi: 10.1088/0031-9155/51/8/011. Epub 2006 Apr 3.
4
Evaluation of the veridose QC phantom.
Med Dosim. 2003 Spring;28(1):49-54. doi: 10.1016/S0958-3947(02)00249-2.
5
Performance characteristics of a gated fiber-optic-coupled dosimeter in high-energy pulsed photon radiation dosimetry.
Appl Radiat Isot. 2010 Feb;68(2):364-9. doi: 10.1016/j.apradiso.2009.10.042. Epub 2009 Nov 1.
7
Photon beam quality variations of a flattening filter free linear accelerator.
Med Phys. 2010 Jan;37(1):49-53. doi: 10.1118/1.3264617.
8
Monte Carlo commissioning of clinical electron beams using large field measurements.
Phys Med Biol. 2010 Jul 21;55(14):4083-105. doi: 10.1088/0031-9155/55/14/009. Epub 2010 Jul 5.
9
Monte Carlo simulation of MOSFET detectors for high-energy photon beams using the PENELOPE code.
Phys Med Biol. 2007 Jan 7;52(1):303-16. doi: 10.1088/0031-9155/52/1/020. Epub 2006 Dec 20.
10
A virtual photon source model of an Elekta linear accelerator with integrated mini MLC for Monte Carlo based IMRT dose calculation.
Phys Med Biol. 2007 Aug 7;52(15):4449-63. doi: 10.1088/0031-9155/52/15/006. Epub 2007 Jun 26.

引用本文的文献

1
Monitoring beam-quality constancy considering uncertainties associated with ionization chambers in Daily QA3 device.
PLoS One. 2021 Feb 17;16(2):e0246845. doi: 10.1371/journal.pone.0246845. eCollection 2021.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验