Department of Biosurgery and Surgical Technology, St Mary's Hospital, Imperial College London, 10th Floor, QEQM Wing, South Wharf Road, London, W2 1NY, UK.
Surg Endosc. 2010 Oct;24(10):2465-74. doi: 10.1007/s00464-010-0987-0. Epub 2010 Mar 31.
Radiofrequency-controlled tissue fusion is a novel technology but the associated lateral thermal damage has not been determined.
Lateral thermal spread of in vivo and ex vivo bowel in a live porcine model fused by radiofrequency energy was evaluated using dynamic infrared thermography and histology.
Mean maximum thermal spread measured on histology was <1.2 mm, with no significant difference between thermal spreads for in vivo and ex vivo bowel for radiofrequency energy delivered at 50 V (p = 0.98) and 100 V (p = 0.85). Mean total maximum thermal spread measured by dynamic infrared thermography was <3.9 mm wide on both sides of the instrument with no significant difference between thermal spreads for in vivo and ex vivo bowel for radiofrequency energy delivered at 50 V (p = 0.34) and 100 V (p = 0.19). Fusion quality for in vivo tissue was better when radiofrequency energy was delivered at 100 V compared with 50 V. However, thermal spread measurements and maximum temperatures reached in the tissue were similar in well- and poorly fused bowel. Thermal changes in well-fused bowel were more uniform throughout the different bowel wall layers, whereas in poorly fused tissues, the mucosa did not show thermally induced changes. There were no significant differences between the maximum temperatures detected for in vivo and ex vivo bowel for radiofrequency energy delivered at 50 V (p = 0.25) and 100 V (p = 0.14).
The total thermal changes at both sides of fused bowel are <3.9 mm. The heat sink effect of the application instrument overshadowed any effects of perfusion on limiting thermal spread. Also, using greater amounts of radiofrequency energy at 100 V to achieve better quality fusion does not necessarily increase lateral thermal damage compared with 50 V.
射频控温热凝是一种新型技术,但相关的侧向热损伤尚未确定。
在活体猪模型中,使用动态红外热成像和组织学评估射频能量融合时体内和体外肠道的侧向热扩散。
组织学上测量的平均最大热扩散<1.2mm,射频能量在 50V(p=0.98)和 100V(p=0.85)时,体内和体外肠道的热扩散无显著差异。动态红外热成像测量的总最大热扩散平均值<仪器两侧 3.9mm,射频能量在 50V(p=0.34)和 100V(p=0.19)时,体内和体外肠道的热扩散无显著差异。与 50V 相比,100V 时体内组织的融合质量更好。然而,在融合良好和融合不良的肠道中,热扩散测量值和达到的最高温度相似。在融合良好的肠道中,热变化在不同的肠壁层之间更加均匀,而在融合不良的组织中,黏膜没有显示出热诱导的变化。50V(p=0.25)和 100V(p=0.14)时,体内和体外肠道检测到的最高温度无显著差异。
融合肠两侧的总热变化<3.9mm。应用仪器的热汇效应掩盖了灌注对限制热扩散的任何影响。此外,与 50V 相比,使用 100V 时更多的射频能量来实现更好的融合质量不一定会增加侧向热损伤。