Edwin R P, Dudermel M T, Lamare M
UK National Physical Laboratory, UK.
Appl Opt. 1982 Mar 1;21(5):878-81. doi: 10.1364/AO.21.000878.
The refractive index of germanium was measured for ten samples, covering the various crystal types, over the 8-14-microm wavelength range. Samples were obtained from several suppliers, and differences between samples in the index values reported were not significant compared with the estimated index uncertainties of +/-0.0003. Three samples were examined at two laboratories to act as a check on the absolute values obtained. A temperature coefficient index of 0.00040 degrees C(-1) was obtained for the 20-25-C range.