屏气自动校准相位对比成像。

Breathheld autocalibrated phase-contrast imaging.

机构信息

Department of Radiology, Stanford University, Stanford, California 94305, USA.

出版信息

J Magn Reson Imaging. 2010 Apr;31(4):1004-14. doi: 10.1002/jmri.22127.

Abstract

PURPOSE

To compare generalized autocalibrating partially parallel acquisitions (GRAPPA), modified sensitivity encoding (mSENSE), and SENSE in phase-contrast magnetic resonance imaging (PC-MRI) applications.

MATERIALS AND METHODS

Aliasing of the torso can occur in PC-MRI applications. If the data are further undersampled for parallel imaging, SENSE can be problematic in correctly unaliasing signals due to coil sensitivity maps that do not match that of the aliased volume. Here, a method for estimating coil sensitivities in flow applications is described. Normal volunteers (n = 5) were scanned on a 1.5 T MRI scanner and underwent PC-MRI scans using GRAPPA, mSENSE, SENSE, and conventional PC-MRI acquisitions. Peak velocity and flow through the aorta and pulmonary artery were evaluated.

RESULTS

Bland-Altman statistics for flow in the aorta and pulmonary artery acquired with mSENSE and GRAPPA methods (R = 2 and R = 3 cases) have comparable mean differences to flow acquired with conventional PC-MRI. GRAPPA and mSENSE PC-MRI have more robust measurements than SENSE when there is aliasing artifact caused by insufficient coil sensitivity maps. For peak velocity, there are no considerable differences among the mSENSE, GRAPPA, and SENSE reconstructions and are comparable to conventional PC-MRI.

CONCLUSION

Flow measurements of images reconstructed with autocalibration techniques have comparable agreement with conventional PC-MRI and provide robust measurements in the presence of wraparound.

摘要

目的

比较广义自动校准部分并行采集(GRAPPA)、改良灵敏度编码(mSENSE)和相位对比磁共振成像(PC-MRI)应用中的灵敏度编码。

材料与方法

PC-MRI 应用中可能会出现体部混叠。如果数据进一步欠采样进行并行成像,由于灵敏度图与混叠体积不匹配,灵敏度编码可能会导致信号无法正确解混叠。本文描述了一种用于在流动应用中估计线圈灵敏度的方法。对 5 名正常志愿者进行了 1.5T MRI 扫描仪扫描,并进行了 GRAPPA、mSENSE、SENSE 和传统 PC-MRI 采集的 PC-MRI 扫描。评估了主动脉和肺动脉的峰值速度和流量。

结果

mSENSE 和 GRAPPA 方法(R = 2 和 R = 3 例)采集的主动脉和肺动脉流量的 Bland-Altman 统计数据与传统 PC-MRI 采集的流量具有可比的平均差异。当线圈灵敏度图不足导致混叠伪影时,GRAPPA 和 mSENSE PC-MRI 的测量比 SENSE 更稳健。对于峰值速度,mSENSE、GRAPPA 和 SENSE 重建之间没有明显差异,与传统 PC-MRI 相当。

结论

使用自动校准技术重建的图像的流量测量与传统 PC-MRI 具有相当的一致性,并在存在缠绕的情况下提供稳健的测量。

相似文献

1
Breathheld autocalibrated phase-contrast imaging.
J Magn Reson Imaging. 2010 Apr;31(4):1004-14. doi: 10.1002/jmri.22127.
3
Comparison of in vitro and in vivo MRI of the spine using parallel imaging.
AJR Am J Roentgenol. 2004 Mar;182(3):749-55. doi: 10.2214/ajr.182.3.1820749.
4
Field-of-view limitations in parallel imaging.
Magn Reson Med. 2004 Nov;52(5):1118-26. doi: 10.1002/mrm.20249.
5
Peak velocity and flow quantification validation for sensitivity-encoded phase-contrast MR imaging.
Acad Radiol. 2007 Mar;14(3):258-69. doi: 10.1016/j.acra.2006.11.008.
9
Accelerated volumetric MRI with a SENSE/GRAPPA combination.
J Magn Reson Imaging. 2006 Aug;24(2):444-50. doi: 10.1002/jmri.20632.

引用本文的文献

1
Parallel MR imaging.
J Magn Reson Imaging. 2012 Jul;36(1):55-72. doi: 10.1002/jmri.23639.
2
Acceleration of tissue phase mapping with sensitivity encoding at 3T.
J Cardiovasc Magn Reson. 2011 Oct 12;13(1):59. doi: 10.1186/1532-429X-13-59.

本文引用的文献

1
Iterative GRAPPA (iGRAPPA) for improved parallel imaging reconstruction.
Magn Reson Med. 2008 Apr;59(4):903-7. doi: 10.1002/mrm.21370.
2
Tailored utilization of acquired k-space points for GRAPPA reconstruction.
J Magn Reson. 2005 May;174(1):60-7. doi: 10.1016/j.jmr.2005.01.015.
3
The SENSE ghost: field-of-view restrictions for SENSE imaging.
J Magn Reson Imaging. 2004 Dec;20(6):1046-51. doi: 10.1002/jmri.20204.
4
Field-of-view limitations in parallel imaging.
Magn Reson Med. 2004 Nov;52(5):1118-26. doi: 10.1002/mrm.20249.
5
UNFOLD-SENSE: a parallel MRI method with self-calibration and artifact suppression.
Magn Reson Med. 2004 Aug;52(2):310-20. doi: 10.1002/mrm.20133.
6
Flow volume and shunt quantification in pediatric congenital heart disease by real-time magnetic resonance velocity mapping: a validation study.
Circulation. 2004 Apr 27;109(16):1987-93. doi: 10.1161/01.CIR.0000126494.66859.A2. Epub 2004 Apr 5.
7
Accuracy and reproducibility in phase contrast imaging using SENSE.
Magn Reson Med. 2003 Nov;50(5):1061-8. doi: 10.1002/mrm.10634.
8
Rapid left-to-right shunt quantification in children by phase-contrast magnetic resonance imaging combined with sensitivity encoding (SENSE).
Circulation. 2003 Sep 16;108(11):1355-61. doi: 10.1161/01.CIR.0000087603.97036.C2. Epub 2003 Aug 25.
9
Generalized autocalibrating partially parallel acquisitions (GRAPPA).
Magn Reson Med. 2002 Jun;47(6):1202-10. doi: 10.1002/mrm.10171.
10
A generalized approach to parallel magnetic resonance imaging.
Med Phys. 2001 Aug;28(8):1629-43. doi: 10.1118/1.1386778.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验