J Magn Reson Imaging. 2010 Apr;31(4):1004-14. doi: 10.1002/jmri.22127.
To compare generalized autocalibrating partially parallel acquisitions (GRAPPA), modified sensitivity encoding (mSENSE), and SENSE in phase-contrast magnetic resonance imaging (PC-MRI) applications.
Aliasing of the torso can occur in PC-MRI applications. If the data are further undersampled for parallel imaging, SENSE can be problematic in correctly unaliasing signals due to coil sensitivity maps that do not match that of the aliased volume. Here, a method for estimating coil sensitivities in flow applications is described. Normal volunteers (n = 5) were scanned on a 1.5 T MRI scanner and underwent PC-MRI scans using GRAPPA, mSENSE, SENSE, and conventional PC-MRI acquisitions. Peak velocity and flow through the aorta and pulmonary artery were evaluated.
Bland-Altman statistics for flow in the aorta and pulmonary artery acquired with mSENSE and GRAPPA methods (R = 2 and R = 3 cases) have comparable mean differences to flow acquired with conventional PC-MRI. GRAPPA and mSENSE PC-MRI have more robust measurements than SENSE when there is aliasing artifact caused by insufficient coil sensitivity maps. For peak velocity, there are no considerable differences among the mSENSE, GRAPPA, and SENSE reconstructions and are comparable to conventional PC-MRI.
Flow measurements of images reconstructed with autocalibration techniques have comparable agreement with conventional PC-MRI and provide robust measurements in the presence of wraparound.
比较广义自动校准部分并行采集(GRAPPA)、改良灵敏度编码(mSENSE)和相位对比磁共振成像(PC-MRI)应用中的灵敏度编码。
PC-MRI 应用中可能会出现体部混叠。如果数据进一步欠采样进行并行成像,由于灵敏度图与混叠体积不匹配,灵敏度编码可能会导致信号无法正确解混叠。本文描述了一种用于在流动应用中估计线圈灵敏度的方法。对 5 名正常志愿者进行了 1.5T MRI 扫描仪扫描,并进行了 GRAPPA、mSENSE、SENSE 和传统 PC-MRI 采集的 PC-MRI 扫描。评估了主动脉和肺动脉的峰值速度和流量。
mSENSE 和 GRAPPA 方法(R = 2 和 R = 3 例)采集的主动脉和肺动脉流量的 Bland-Altman 统计数据与传统 PC-MRI 采集的流量具有可比的平均差异。当线圈灵敏度图不足导致混叠伪影时,GRAPPA 和 mSENSE PC-MRI 的测量比 SENSE 更稳健。对于峰值速度,mSENSE、GRAPPA 和 SENSE 重建之间没有明显差异,与传统 PC-MRI 相当。
使用自动校准技术重建的图像的流量测量与传统 PC-MRI 具有相当的一致性,并在存在缠绕的情况下提供稳健的测量。