A clinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500 (Japan) was performed to evaluate validity and repeatability compared with non-cycloplegic subjective refraction and Javal-Schiotz keratometry. An investigation into the dynamic recording capabilities of the instrument was also conducted.
Refractive error measurements were obtained from 150 eyes of 75 subjects (aged 25.12 +/- 9.03 years), subjectively by a masked optometrist, and objectively with the WAM-5500 at a second session. Keratometry measurements from the WAM-5500 were compared to Javal-Schiotz readings. Intratest variability was examined on all subjects, whilst intertest variability was assessed on a subgroup of 44 eyes 7-14 days after the initial objective measures. The accuracy of the dynamic recording mode of the instrument and its tolerance to longitudinal movement was evaluated using a model eye. An additional evaluation of the dynamic mode was performed using a human eye in relaxed and accommodated states.
Refractive error determined by the WAM-5500 was found to be very similar (p = 0.77) to subjective refraction (difference, -0.01 +/- 0.38 D). The instrument was accurate and reliable over a wide range of refractive errors (-6.38 to +4.88 D). WAM-5500 keratometry values were steeper by approximately 0.05 mm in both the vertical and horizontal meridians. High intertest repeatability was demonstrated for all parameters measured: for sphere, cylinder power and MSE, over 90% of retest values fell within +/-0.50 D of initial testing. In dynamic (high-speed) mode, the root-mean-square of the fluctuations was 0.005 +/- 0.0005 D and a high level of recording accuracy was maintained when the measurement ring was significantly blurred by longitudinal movement of the instrument head.
The WAM-5500 Auto Ref/Keratometer represents a reliable and valid objective refraction tool for general optometric practice, with important additional features allowing pupil size determination and easy conversion into high-speed mode, increasing its usefulness post-surgically following accommodating intra-ocular lens implantation, and as a research tool in the study of accommodation.
对 Grand Seiko Auto Ref/Keratometer WAM-5500(日本)进行临床评估,以评估其与非睫状肌麻痹主观验光和 Javal-Schiotz 角膜曲率计相比的有效性和可重复性。还对仪器的动态记录能力进行了调查。
对 75 名受试者(年龄 25.12 ± 9.03 岁)的 150 只眼睛进行了屈光误差测量,由一名掩蔽验光师进行主观测量,然后在第二次就诊时使用 WAM-5500 进行客观测量。将 WAM-5500 的角膜曲率计测量值与 Javal-Schiotz 读数进行比较。对所有受试者进行了日内可变性检查,对 44 只眼睛进行了 7-14 天的初始客观测量后的组间可变性评估。使用模型眼评估仪器的动态记录模式的准确性及其对纵向运动的容差。使用放松和调节状态下的人眼对动态模式进行了额外评估。
发现 WAM-5500 确定的屈光误差与主观验光非常相似(p = 0.77)(差值,-0.01 ± 0.38 D)。该仪器在广泛的屈光不正范围内(-6.38 至 +4.88 D)均准确可靠。WAM-5500 角膜曲率计值在垂直和水平子午线方向上分别约陡峭 0.05 毫米。所有测量参数均表现出高度的组间可重复性:对于球镜、柱镜力和均方根误差,超过 90%的复测值落在初始测试值的±0.50 D 以内。在动态(高速)模式下,波动的均方根为 0.005 ± 0.0005 D,当仪器头部的纵向运动显著模糊测量环时,仍保持较高的记录精度。
WAM-5500 自动 Ref/Keratometer 是一种可靠且有效的客观验光工具,适用于一般验光实践,具有重要的附加功能,可确定瞳孔大小,并轻松转换为高速模式,在植入调节性人工晶状体后术后增加其有用性,并作为研究工具在调节研究中。