BACKGROUND

Mammography has a lower sensitivity for breast cancer detection in younger women and those with dense breasts. Recent improvements in digital infrared breast imaging suggest there may be a role for this technology and we have studied its performance in 100 women prior to breast needle core biopsy (CB).

METHODS

All patients were imaged using a digital infrared breast (DIB) scan (Sentinel BreastScan) prior to breast biopsy. Analysis of the infrared scans was performed, blinded to biopsy results, in four different ways: Sentinel screening report, Sentinel artificial intelligence (neural network), expert manual review and NoTouch BreastScan a novel artificial intelligence programme.

RESULTS

Of 106 biopsies performed in 100 women, 65 were malignant and 41 were benign. Sensitivity of Sentinel screening (53%) and Sentinel neural network (48%) was low but analysis with NoTouch software (70%) was much closer to expert manual review (78%). Sensitivity (78%) and specificity (75%) using NoTouch BreastScan were higher in women under 50 and the combination of mammography and DIB, with NoTouch interpretation, in this age group resulted in a sensitivity of 89%.

CONCLUSION

DIB using NoTouch is an effective adjunctive test for breast cancer detection in women under 70 and appears to be particularly effective in women under 50 where maximal sensitivity (78%) and specificity (75%) were observed. The combined sensitivity of NoTouch BreastScan and mammography in women under 50 was encouraging at 89%, suggesting a potential way forward for a dual imaging approach in this younger age group.

摘要

背景

在年轻女性和乳腺密度较高的女性中,乳房 X 光摄影术对乳腺癌的检测灵敏度较低。最近数字红外乳腺成像技术的改进表明,该技术可能具有一定的作用,我们在 100 例乳腺针芯活检(CB)前患者中研究了该技术的性能。

方法

所有患者均在乳腺活检前使用数字红外乳腺(DIB)扫描(Sentinel BreastScan)进行成像。对红外扫描进行分析,分析结果不考虑活检结果,采用四种不同方法进行:Sentinel 筛查报告、Sentinel 人工智能(神经网络)、专家手动审查和 NoTouch BreastScan 新型人工智能程序。

结果

在 100 例女性的 106 例活检中,65 例为恶性,41 例为良性。Sentinel 筛查(53%)和 Sentinel 神经网络(48%)的灵敏度较低,但 NoTouch 软件分析(70%)与专家手动审查(78%)更接近。NoTouch BreastScan 在 50 岁以下女性中的灵敏度(78%)和特异性(75%)更高,在该年龄组中,联合使用乳房 X 光摄影术和 DIB,以及 NoTouch 解释,其灵敏度为 89%。

结论

NoTouch 红外乳腺成像术是一种有效的辅助检测 70 岁以下女性乳腺癌的方法,在 50 岁以下女性中尤其有效,观察到的最大灵敏度(78%)和特异性(75%)。在 50 岁以下女性中,NoTouch BreastScan 和乳房 X 光摄影术的联合灵敏度为 89%,令人鼓舞,这表明在该年轻年龄组中,双成像方法可能具有一定的应用前景。