光纤功率计的校准:连接器的影响。

Calibration of optical fiber power meters: the effect of connectors.

作者信息

Gallawa R L, Li X

出版信息

Appl Opt. 1987 Apr 1;26(7):1170-4. doi: 10.1364/AO.26.001170.

Abstract

This paper addresses the question of accurate measurement of optical power at the wavelengths and power levels of interest to the telecommunications community. In particular, we examine the calibration of power meters that are destined for use in a field environment. Connectors and adapters are shown to skew the measurements, leading to errors attributable to reflections from the connector or to angular dependence of detector response. Calibration data are taken using two popular connector types: a biconic and an SMA type. The data are sufficient to illustrate the problem but definitive conclusions cannot be drawn regarding variability of performance with connector or connector type, because of the limited data.

摘要

本文探讨了在电信领域感兴趣的波长和功率水平下准确测量光功率的问题。特别是,我们研究了用于现场环境的功率计的校准。结果表明,连接器和适配器会使测量结果产生偏差,导致误差归因于连接器的反射或探测器响应的角度依赖性。使用两种常见的连接器类型(双锥型和SMA型)获取校准数据。这些数据足以说明问题,但由于数据有限,无法就不同连接器或连接器类型的性能变化得出明确结论。