Inadequate gas supply to patients with an adjustable pressure-limiting valve in the fully opened position.

机构信息

Department of Anesthesiology, Hachioji Medical Center, Tokyo Medical University, 1163 Tatemachi, Hachioji, Tokyo, 193-0998, Japan.

出版信息

J Anesth. 2010 Aug;24(4):626-9. doi: 10.1007/s00540-010-0958-6. Epub 2010 May 13.

Abstract

With spontaneous ventilation, sufficient exhaust resistance at the adjustable pressure-limiting (APL) valve when fully opened ensures that the reservoir bag fills and adequately supplies gas to patients. A lack of exhaust resistance with the APL valve fully open caused inadequate gas supply to patients with four types of anesthesia machines: SA2 (Dräger), Excel-210 SE (GE), Fabius (Dräger), and Cato (Dräger). Mechanically, the SA2 and Excel-210 SE APL valve systems, which are of the spring-loaded disc type positioned horizontally, cannot maintain sufficient exhaust resistance with the APL valve fully open. As for the Fabius and Cato, an exhaust valve independent of the APL valve should maintain sufficient exhaust resistance continuously. However, accumulated viscous substances on the thin diaphragm of the exhaust valve contributed to hindrance of diaphragm closure.

摘要

在自主通气时,当可调压力限制(APL)阀完全打开时,足够的排气阻力可确保储气囊充满并向患者提供充足的气体。当 APL 阀完全打开时,如果没有排气阻力,四种麻醉机(SA2(德尔格)、Excel-210 SE(GE)、Fabius(德尔格)和 Cato(德尔格))将无法向患者提供充足的气体。从机械角度来看,SA2 和 Excel-210 SE APL 阀系统采用水平放置的弹簧加载盘式,当 APL 阀完全打开时,无法保持足够的排气阻力。对于 Fabius 和 Cato,APL 阀独立的排气阀应持续保持足够的排气阻力。然而,排气阀薄膜上积聚的粘性物质会阻碍膜片关闭。