Kojima Osamu, Kuzukawa Akiko, Sai Yosikazu
Department of Anesthesiology, Toyosato Hospital, Shiga 529-1168.
Masui. 2010 Nov;59(11):1405-7.
We experienced inadequate gas supply to the patients from the breathing circuit of anesthesia delivery system with APL valve at full open position. It was hardly noticed after the check-up before starting work with anesthesia delivery system. This problem was caused by the mistake in installing the one-way valve when the repair of anesthesia delivery system was done. In addition to the check-up before starting work, we should check whether gas comes out from the breathing circuit.
我们发现,在APL阀完全打开的情况下,麻醉输送系统的呼吸回路向患者供应的气体不足。在开始使用麻醉输送系统之前进行检查时,这个问题几乎没有被发现。这个问题是在麻醉输送系统维修时单向阀安装错误造成的。除了在开始工作前进行检查外,我们还应检查呼吸回路是否有气体流出。