[麻醉输送系统修复后患者气体供应不足]

[Inadequate gas supply to the patients after the repair of anesthesia delivery system].

作者信息

Kojima Osamu, Kuzukawa Akiko, Sai Yosikazu

机构信息

Department of Anesthesiology, Toyosato Hospital, Shiga 529-1168.

出版信息

Masui. 2010 Nov;59(11):1405-7.

Abstract

We experienced inadequate gas supply to the patients from the breathing circuit of anesthesia delivery system with APL valve at full open position. It was hardly noticed after the check-up before starting work with anesthesia delivery system. This problem was caused by the mistake in installing the one-way valve when the repair of anesthesia delivery system was done. In addition to the check-up before starting work, we should check whether gas comes out from the breathing circuit.

摘要

我们发现,在APL阀完全打开的情况下,麻醉输送系统的呼吸回路向患者供应的气体不足。在开始使用麻醉输送系统之前进行检查时,这个问题几乎没有被发现。这个问题是在麻醉输送系统维修时单向阀安装错误造成的。除了在开始工作前进行检查外,我们还应检查呼吸回路是否有气体流出。