Lu D F, Joseph P M
Department of Radiology, University of Pennsylvania, Philadelphia 19104.
Magn Reson Med. 1991 May;19(1):180-5. doi: 10.1002/mrm.1910190117.
A technique was developed to change the resonant frequency of a rotating field birdcage coil from 19F resonance (76.14 MHz) to 1H resonance (80.92 MHz). This was done by inserting an additional shield with a smaller radius on the coil which resonates at 19F. It was shown that the inhomogeneity introduced by reducing the shield radius is not significant. A working coil built with this technique produced both 19F and 1H images satisfactorily.
开发了一种技术,可将旋转磁场鸟笼式线圈的共振频率从19F共振(76.14兆赫兹)改变为1H共振(80.92兆赫兹)。这是通过在19F共振的线圈上插入一个半径较小的附加屏蔽来实现的。结果表明,减小屏蔽半径所引入的不均匀性并不显著。用这种技术制造的工作线圈能够令人满意地产生19F和1H图像。