BACKGROUND

Until recently, Canadian Blood Services (CBS) was performing quality control sterility testing of blood components using three different processes. This study was conducted in order to standardize sterility testing at all CBS centers in a cost-effective manner using the BacT/ALERT 3D system.

METHODS

Blood components including fresh frozen plasma, platelet concentrates, and red blood cells were inoculated with eight bacterial species at target concentrations of 1 and 10 CFU/mL. Pre- and post-spiked samples were inoculated into BacT/ALERT aerobic and anaerobic culture bottles and incubated for a maximum of 10 days. Specificity of the positive culture bottles was verified by Gram staining. Positive results obtained pre- and post-implementation of the in-house sterility testing program at CBS were collected and analyzed.

RESULTS

The BacT/ALERT3D system detected all bacteria in all blood components tested in this validation. Positive cultures were obtained within 28 h of incubation with the exception of Propionibacterium acnes which was detected within 134 h. The percentage of positive cultures ranged from 0.01% to 0.2%. All contaminants isolated were either normal skin flora or environmental microorganisms.

CONCLUSIONS

This study demonstrates the capability of the BacT/ALERT3D system to detect aerobic and anaerobic bacterial contamination in all tested blood components, thereby supporting its use for quality control sterility testing and not only bacterial screening. A standardized process will allow CBS to evaluate and compare blood collection and manufacturing practices across the country.

摘要

背景

直到最近,加拿大血液服务中心(CBS)一直使用三种不同的方法来对血液成分进行质量控制无菌检测。这项研究旨在通过使用 BacT/ALERT 3D 系统以具有成本效益的方式在所有 CBS 中心标准化无菌检测。

方法

将包括新鲜冷冻血浆、血小板浓缩物和红细胞在内的血液成分接种八种细菌,目标浓度为 1 和 10 CFU/mL。预接种和后接种的样本接种到 BacT/ALERT 有氧和厌氧培养瓶中,培养时间最长为 10 天。阳性培养瓶的特异性通过革兰氏染色进行验证。收集并分析了 CBS 内部无菌检测程序实施前后获得的阳性结果。

结果

在这项验证中,BacT/ALERT3D 系统检测到了所有测试血液成分中的所有细菌。除痤疮丙酸杆菌外,所有细菌在培养 28 小时内均得到阳性结果,而痤疮丙酸杆菌则在 134 小时内得到阳性结果。阳性培养的百分比从 0.01%到 0.2%不等。所有分离出的污染物均为正常皮肤菌群或环境微生物。

结论

这项研究表明 BacT/ALERT3D 系统能够检测所有测试血液成分中的需氧和厌氧细菌污染,从而支持其用于质量控制无菌检测,而不仅仅是细菌筛查。标准化的过程将允许 CBS 评估和比较全国的血液采集和制造实践。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验