Liu Tie-long, Yan Wang-jun, Han Yu, Ye Xiao-jian, Jia Lian-shun, Li Jia-shun, Yuan Wen
General Hospital of PLA, Beijing 100853, China.
Nan Fang Yi Ke Da Xue Xue Bao. 2010 May;30(5):1127-31.
To compare the biomechanical performances of different wires and cable fixation devices in posterior instrumentation for atlantoaxial instability, and test the effect of different fixation strengths and fixation approaches on the surgical outcomes.
Six specimens of the atlantoaxial complex (C0-C3) were used to establish models of the normal complex, unstable complex (type II odontoid fracture) and fixed complex. On the wd-5 mechanical testing machine, the parameters including the strength and rigidity of anti-rotation, change and strength of stress, and stability were measured for the normal complex, atlantoaxial instability complex, the new type titanium cable fixation system, Atlas titanium cable, Songer titanium cable, and stainless wire.
The strength and rigidity of anti-rotation, change and strength of stress, stability of flexion, extension and lateral bending of the unstable atlantoaxial complex fixed by the new double locking titanium cable fixation system were superior to those of the Songer or Atlas titanium cable (P<0.05) and medical stainless wire (P<0.05). Simultaneous cable fastening on both sides resulted in better fixation effect than successive cable fastening (P<0.05). Better fixation effect was achieved by fastening the specimen following a rest (P<0.05).
The fixation effects can be enhanced by increased fastening strengths. The new type double locking titanium cable fixation system has better biomechanical performance than the conventional Songer and Atlas titanium cables. Fastening the unstable specimens after a rest following simultaneous fastening of the specimen on both sides produces better fixation effect.
比较不同钢丝和缆线固定装置在寰枢椎不稳后路内固定中的生物力学性能,测试不同固定强度和固定方式对手术效果的影响。
使用6个寰枢椎复合体(C0-C3)标本建立正常复合体、不稳定复合体(Ⅱ型齿状突骨折)和固定复合体模型。在WD-5力学试验机上,测量正常复合体、寰枢椎不稳复合体、新型钛缆固定系统、寰椎钛缆、宋格钛缆和不锈钢丝的抗旋转强度和刚度、应力变化和强度以及稳定性等参数。
新型双锁定钛缆固定系统固定的不稳定寰枢椎复合体在抗旋转强度和刚度、应力变化和强度、屈伸及侧方弯曲稳定性方面均优于宋格或寰椎钛缆(P<0.05)和医用不锈钢丝(P<0.05)。双侧同时缆线固定的固定效果优于先后缆线固定(P<0.05)。标本休息后固定的固定效果更好(P<0.05)。
增加固定强度可提高固定效果。新型双锁定钛缆固定系统的生物力学性能优于传统的宋格和寰椎钛缆。双侧同时固定标本后休息再固定不稳定标本可产生更好的固定效果。