Korean J Anesthesiol. 2010 May;58(5):480-4. doi: 10.4097/kjae.2010.58.5.480. Epub 2010 May 31.
Transesophageal echocardiography (TEE) has an important role during congenital heart surgery. TEE in small infants is associated with complications, including an inability to pass the TEE probe, esophageal trauma, airway compression, aortic compression, and interference with ventilation. Recently, a monoplane intravascular ultrasound catheter (IVUC) has been developed for intracardiac echocardiography. The efficacy of IVUC for transesophageal use has been shown in numerous animal studies and several human studies, but there have been few reports involving small infants using an IVUC probe. We examined 15 pediatric patients undergoing congenital cardiac surgery using an 8-Fr AcuNav IVUC probe. We checked the cardiac anatomy, cardiac function pre-operatively and de-aeration before weaning from CPB; the surgical repairs were evaluated post-operatively. Although the IVUC probe has limitations associated with the monoplane, we found the IVUC probe to be useful in small infants and safer than the TEE.
经食管超声心动图(TEE)在先天性心脏病手术中具有重要作用。在小婴儿中进行 TEE 检查会引起一些并发症,包括 TEE 探头无法通过、食管损伤、气道压迫、主动脉压迫和干扰通气等。最近,一种单平面血管内超声导管(IVUC)已被开发用于心脏内超声检查。大量动物研究和几项人体研究已经证明了 IVUC 在经食管使用的有效性,但涉及使用 IVUC 探头的小婴儿的报道很少。我们使用 8Fr AcuNav IVUC 探头检查了 15 例接受先天性心脏手术的儿科患者。我们在 CPB 脱机前检查了心脏解剖结构、术前心功能和脱气情况;并在术后评估了手术修复情况。尽管 IVUC 探头存在与单平面相关的局限性,但我们发现 IVUC 探头在小婴儿中很有用,而且比 TEE 更安全。