使用毛细管电泳-非接触电导检测技术快速监测临床样本中的氨基酸阵列。

Rapid monitoring of arrays of amino acids in clinical samples using capillary electrophoresis with contactless conductivity detection.

机构信息

Institute of Biochemistry, Cell and Molecular Biology, Third Faculty of Medicine, Charles University, Czech Republic.

出版信息

J Sep Sci. 2010 Aug;33(16):2394-401. doi: 10.1002/jssc.201000137.

Abstract

CE with contactless conductivity detection has been used to separate 28 biogenic amino acids in a short capillary with an effective length of 18 cm. All the tested amino acids can be mutually separated in 0.5-10 mol/L acetic acid electrolytes. The time of analysis does not exceed 6 min and the LODs vary from 0.1 to 1.7 micromol/L. The CVs lie within the intervals 0.01-0.4% and 0.9-4% for the migration times and the analyte peak areas, respectively. The procedure has been successfully applied to the determinations of the whole amino acid spectra in blood plasma, urine, saliva and cerebrospinal fluid samples.

摘要

无接触式电导检测的 CE 已被用于在有效长度为 18 厘米的短毛细管中分离 28 种生物衍生氨基酸。在 0.5-10 mol/L 乙酸电解质中,所有测试的氨基酸都可以相互分离。分析时间不超过 6 分钟,LOD 范围为 0.1 至 1.7 微摩尔/升。对于迁移时间和分析物峰面积,CV 分别在 0.01-0.4%和 0.9-4%的区间内。该方法已成功应用于测定血浆、尿液、唾液和脑脊液样品中的全氨基酸谱。