Anwar Eman M, Kheiralla Lamia S, Tammam Riham H
Fixed Prosthodontic Department, Cairo University, Cairo, Egypt.
J Oral Implantol. 2011 Jun;37(3):309-17. doi: 10.1563/AAID-JOI-D-09-00084. Epub 2010 Jun 14.
The effect of fluoride ion concentration on the corrosion behavior of Ti and Ti6Al4V implant alloys, when coupled with either metal/ceramic or all-ceramic superstructure, was examined by different electrochemical methods in artificial saliva solutions. It was concluded that increased fluoride concentration leads to a decrease in the corrosion resistance of all tested couples. The type of the superstructure also showed a significant effect on the corrosion resistance of the couple.
在人工唾液溶液中,采用不同的电化学方法研究了氟离子浓度对与金属/陶瓷或全陶瓷上部结构耦合的钛及Ti6Al4V植入合金腐蚀行为的影响。得出的结论是,氟化物浓度的增加会导致所有测试组合的耐腐蚀性降低。上部结构的类型对组合的耐腐蚀性也有显著影响。