Watanabe Ichiro, Nuo Min, Okumura Yasuo, Ohkubo Kimie, Ashino Sonoko, Kofune Masayoshi, Kofune Tatsuya, Nakai Toshiko, Kasamaki Yuji, Hirayama Atsushi
Division of Cardiology, Department of Medicine, Nihon University School of Medicine, 30-1 Oyaguchi-kami, Itabashi-ku, Tokyo 173-8610, Japan.
Int Heart J. 2010 May;51(3):193-8. doi: 10.1536/ihj.51.193.
Steam pop and intramural charring have been reported during cooled-tip radiofrequency catheter ablation (RFCA). We studied the feasibility of temperature-controlled cooled-tip RFCA in the canine heart.An internally cooled ablation catheter was inserted into the left ventricle. A custom-made radiofrequency (RF) generator capable of controlling the tip-temperature at the preset level by slow increases in the power was used. Temperature-controlled cooled-tip RF applications were performed at a target temperature of 40 degrees C for 90 seconds. Acute study: Intramyocardial temperature was measured at the ablation site in 10 dogs by inserting a fluoroptic probe. Chronic study: Lesion depth and volume were measured in 5 dogs after 3 weeks of survival. In the acute study, no pop or abrupt impedance rise was observed. Maximum intramyocardial temperature was 72.4 + or - 14.4 degrees C at 2-4 mm above the endocardium. No coagulum formation, craters, or intramural charring were observed. Maximum lesion depth was 6.7 + or - 1.5 mm, and lesion volume was 404 + or - 219 mm3. In the chronic study, maximum lesion depth was 5.9 + or - 1.1 mm, and lesion volume was 281 + or - 210 mm(3).Temperature controlled RFCA is feasible with a cooled-tip catheter and an RF generator that slowly increases the RF power until the preset catheter-tip temperature is reached.