CSTAR, Lawson Health Research Institute, 339 Windermere Road, London, ON, N6A 5A5, Canada.
Surg Endosc. 2011 Jan;25(1):186-92. doi: 10.1007/s00464-010-1155-2. Epub 2010 Jun 18.
Natural orifice transluminal endoscopic surgery (NOTES) may represent the next frontier for therapeutic minimally invasive surgery; however, its feasibility is currently limited by the lack of suitable instruments. Identifying the forces required to manipulate tissue during NOTES is a necessary first step in the development of better instrumentation.
Sensorized instruments were used to measure the forces acting at the tip of the instruments during transgastric and transperineal NOTES procedures performed in two female pigs. The maximum and average forces when handling tissue were determined and compared.
The results show that, for the transgastric approach, the average forces required are significantly less than in the transperineal approach (43% less), and that the maximum forces required are almost 8 and 16 N in the transgastric and transperineal approaches, respectively. The forces were higher than 5 N in 1.6% of the measurements in the transgastric approach and 2.9% in the transperineal approach.
This study presents an experimental measurement of tissue manipulation forces in a NOTES procedure. This information may be valuable for research groups interested in developing NOTES instruments and devices. It is recommended that NOTES instruments be designed to easily handle forces as high as 16 N.
自然腔道内镜手术(NOTES)可能代表了治疗性微创外科的下一个前沿领域;然而,由于缺乏合适的器械,其可行性目前受到限制。确定在NOTES 中进行组织操作所需的力是开发更好器械的必要的第一步。
使用传感器仪器测量在两只雌性猪中进行的经胃和经会阴 NOTES 手术中仪器尖端的作用力。确定并比较处理组织时的最大和平均力。
结果表明,对于经胃入路,所需的平均力明显小于经会阴入路(少 43%),并且经胃和经会阴入路所需的最大力分别约为 8 和 16 N。在经胃入路中,1.6%的测量值中的力高于 5 N,而在经会阴入路中,2.9%的测量值中的力高于 5 N。
本研究对 NOTES 手术中的组织操作力进行了实验测量。这些信息对于有兴趣开发 NOTES 器械和设备的研究小组可能具有重要价值。建议将 NOTES 仪器设计为能够轻松处理高达 16 N 的力。