磁共振成像术对 Ex-PRESS 不锈钢青光眼引流装置的评估。

MRI of the Ex-PRESS stainless steel glaucoma drainage device.

机构信息

Department of Ophthalmology, Rocky Mountain Lions Eye Institute, University of Colorado at Denver, Aurora, CO 80045, USA.

出版信息

Br J Ophthalmol. 2011 Feb;95(2):251-4. doi: 10.1136/bjo.2009.173906. Epub 2010 Jun 24.

Abstract

AIM

To evaluate the magnetic properties of the Ex-PRESS stainless steel glaucoma drainage device during MRI.

DESIGN

Experimental study.

METHODS

The Ex-PRESS glaucoma drainage device (316L stainless steel) was examined for magnetic field interactions under standard 1.5, 3.0, and 4.7 T MRI scanning protocols. Testing included measurements of translational and rotational motion of the device induced by static magnetic fields. In addition, the change in the temperature of the device was measured to assess the presence of radiofrequency heating of the stainless steel device.

MAIN OUTCOME MEASURES

Degree of angular deflection, device displacement and rotation, and change in temperature.

RESULTS

During induced torque testing, displacement did not occur under 1.5 and 3.0 T conditions, although a significant amount of displacement occurred in the 4.7 T environment. Increasing amounts of angular deflection were demonstrated at all three field strengths. We did not record significant temperature changes during brain MRI sequences at any of the three MRI strengths.

CONCLUSIONS

The Ex-PRESS glaucoma drainage device, manufactured from grade 316L stainless steel, does move in the presence of high magnetic fields. The clinical significance of this finding and translation to in vivo conditions are not currently known. Further studies are needed to better understand how patients might be affected by the magnetic properties of this device post implantation and how patients should be counselled in regards to safety of MRI in the early and late postoperative period.

摘要

目的

评估 Ex-PRESS 不锈钢青光眼引流装置在 MRI 检查中的磁学特性。

设计

实验研究。

方法

对 Ex-PRESS 青光眼引流装置(316L 不锈钢)进行检测,评估其在标准 1.5、3.0 和 4.7 T MRI 扫描方案下的磁场相互作用。测试包括测量由静磁场引起的装置的平移和旋转运动。此外,还测量了装置温度的变化,以评估不锈钢装置是否存在射频加热。

主要观察指标

角偏转程度、装置位移和旋转以及温度变化。

结果

在感应扭矩测试中,1.5 和 3.0 T 条件下未发生位移,而在 4.7 T 环境中发生了大量位移。在所有三种场强下,均表现出越来越大的角偏转。在三种 MRI 强度下,进行脑 MRI 序列时,未记录到明显的温度变化。

结论

由 316L 不锈钢制成的 Ex-PRESS 青光眼引流装置在强磁场中会移动。目前尚不清楚这一发现的临床意义及其在体内条件下的转化。需要进一步研究以更好地了解患者在植入后可能会受到该装置磁性的影响,以及患者在术后早期和晚期应如何就 MRI 安全性进行咨询。