Suppr超能文献

早产儿视网膜病变中视网膜血管宽度和迂曲度的计算机辅助测量

Computer-assisted measurement of retinal vascular width and tortuosity in retinopathy of prematurity.

作者信息

Kiely Amanda E, Wallace David K, Freedman Sharon F, Zhao Zheen

机构信息

School of Medicine, Department of Ophthalmology, Duke University Medical Center, Durham, NC 27710, USA.

出版信息

Arch Ophthalmol. 2010 Jul;128(7):847-52. doi: 10.1001/archophthalmol.2010.133.

Abstract

OBJECTIVES

To validate the accuracy of ROPtool software in measuring retinal vascular width and tortuosity in a large image set compared with expert diagnoses.

METHODS

Tortuosity and dilation indexes generated by ROPtool were compared with 3 expert consensus grades of normal, pre-plus, or plus disease for 368 quadrants in 92 RetCam (Clarity Medical Systems, Pleasanton, California) fundus images. Sensitivity and specificity of ROPtool software in diagnosing tortuosity and dilation sufficient for plus and pre-plus disease were calculated. These measures were compared with individual accuracies of 3 experienced pediatric ophthalmologists.

RESULTS

The mean tortuosity indexes for expert-diagnosed categories of normal, pre-plus, and plus disease were 7.04, 18.73, and 34.62, respectively (P < .001), and the mean dilation indexes were 9.63, 12.05, and 13.61, respectively (P < .001). When optimal tortuosity and dilation index thresholds (from receiver operating characteristic curves) were applied, resultant sensitivity and specificity were 0.913 and 0.863, respectively, for plus tortuosity and 0.782 and 0.840, respectively, for plus dilation. These values were comparable to the performance of examiners judged against the same expert panel.

CONCLUSION

ROPtool version 2.1.5 accurately measures tortuosity and dilation of posterior pole blood vessels in RetCam images, corresponding well with expert diagnostic categories of normal, pre-plus, and plus disease and performing comparably to experienced examiners.

摘要

目的

与专家诊断结果相比,验证ROPtool软件在测量大量图像集中视网膜血管宽度和迂曲度方面的准确性。

方法

将ROPtool生成的迂曲度和扩张指数与92张RetCam(Clarity Medical Systems,普莱森顿,加利福尼亚)眼底图像中368个象限的正常、预加或加病变的3个专家共识等级进行比较。计算ROPtool软件在诊断足以诊断加病变和预加病变的迂曲度和扩张方面的敏感性和特异性。将这些指标与3位经验丰富的儿科眼科医生的个体准确性进行比较。

结果

专家诊断的正常、预加和加病变类别的平均迂曲度指数分别为7.04、18.73和34.62(P <.001),平均扩张指数分别为9.63、12.05和13.61(P <.001)。当应用最佳迂曲度和扩张指数阈值(来自受试者工作特征曲线)时,加迂曲度的敏感性和特异性分别为0.913和0.863,加扩张的敏感性和特异性分别为0.782和0.840。这些值与根据同一专家小组判断的检查者的表现相当。

结论

ROPtool 2.1.5版本能够准确测量RetCam图像中后极血管的迂曲度和扩张度,与正常、预加和加病变的专家诊断类别高度相符,并且与经验丰富的检查者表现相当。