Accuracy of measuring the cortical bone thickness adjacent to dental implants using cone beam computed tomography.

机构信息

Department of Restorative Dentistry, King's College London Dental Institute, London, UK.

出版信息

Clin Oral Implants Res. 2010 Jul;21(7):718-25. doi: 10.1111/j.1600-0501.2009.01905.x.

Abstract

OBJECTIVE

To assess the accuracy of measuring the cortical bone thickness adjacent to dental implants using two cone beam computed tomography (CBCT) systems.

MATERIAL AND METHODS

Ten 4 x 11 mm Astra Tech implants were placed at varying distances from the cortical bone in two prepared bovine ribs. Both ribs were scanned in a reproducible position using two different CBCT scanners. Ten examiners each carried out four measurements on all 10 implants using the two CBCT systems: vertical distance between the top of the implant and the alveolar crest (IT-AC), and thickness of the cortical bone from the outer surface of the implant threads at 3, 6 and 9 mm from the top of the implant. Ground sections were prepared and bone thickness was measured using a light microscope and a graticule to give a gold standard (GS) measurement.

RESULTS

The examiner's measurements were significantly different between CBCT systems for the vertical and thickness dimensions (P<0.001) while measuring the cortical bone thickness between 0.3 and 3.7 mm. Within that range, i-CAT NG measurements were consistently underestimated in comparison with the GS. Accuitomo 3D60 FPD measurements closely approximated the GS, except when cortical bone thickness was <0.8 mm. The mean percentage errors from the GS at 3, 6 and 9 mm measurement levels were 68%, 28% and 18%, respectively, for i-CAT NG and 23%, 5% and 6%, respectively, for Accuitomo 3D60 FPD.

CONCLUSIONS

Within the limitations of this study, it was concluded that i-CAT NG (voxel size 0.3) may not produce sufficient resolution of the thin cortical bone adjacent to dental implants and, therefore, the measurements may not be accurate; whereas, Accuitomo 3D60 FPD (voxel size 0.125) may produce better resolution and more accurate measurement of the thin bone.

摘要

目的

评估两种锥形束计算机断层扫描(CBCT)系统测量牙种植体周围皮质骨厚度的准确性。

材料和方法

将 10 个 4x11mm AstraTech 种植体放置在两个准备好的牛肋骨中,距离皮质骨的不同距离处。两根肋骨以可重复的位置使用两种不同的 CBCT 扫描仪进行扫描。十位检查者每人在两个 CBCT 系统上对所有 10 个种植体进行了四次测量:种植体顶部与牙槽嵴之间的垂直距离(IT-AC),以及从种植体顶部 3、6 和 9mm 处种植体螺纹外表面开始的皮质骨厚度。制备磨片并使用光学显微镜和网格测量骨厚度,以获得金标准(GS)测量值。

结果

检查者的测量值在 CBCT 系统之间存在显著差异,垂直和厚度尺寸(P<0.001),而在测量 0.3 至 3.7mm 之间的皮质骨厚度时。在该范围内,与 GS 相比,i-CAT NG 的测量值始终被低估。Accuitomo 3D60 FPD 的测量值与 GS 非常接近,除了皮质骨厚度<0.8mm 时。在 3、6 和 9mm 测量水平,i-CAT NG 的 GS 平均百分比误差分别为 68%、28%和 18%,Accuitomo 3D60 FPD 的 GS 平均百分比误差分别为 23%、5%和 6%。

结论

在本研究的限制范围内,得出结论,i-CAT NG(体素大小 0.3)可能无法对牙种植体周围的薄皮质骨产生足够的分辨率,因此测量值可能不准确;而 Accuitomo 3D60 FPD(体素大小 0.125)可能会产生更好的分辨率和更准确的薄骨测量值。