Departments of Neurosurgery and Neuroscience, University of Minnesota, Minneapolis, Minnesota 55455, USA.
Neurosurgery. 2010 Aug;67(2):390-7. doi: 10.1227/01.NEU.0000372091.64824.63.
In the United States, the most commonly used surgical treatment for patients with Parkinson's disease is the implantation of deep brain stimulation (DBS) electrodes within the subthalamic nucleus. However, DBS device programming remains difficult and is a possible source of decreased efficacy.
We investigated the relationship between local field potential (LFP) activities in the subthalamic nucleus and the therapeutic response to programming.
We recorded LFPs with macroelectrodes placed unilaterally for DBS in 4 PD patients, 3 weeks after implantation, before the start of log-term DBS. Power-frequency spectra were calculated for each of 7 possible electrode contacts or contact pairs, over multiple 5- to 10-minute quiet waking epochs and over 30-second epochs during hand movements. Subsequently, DBS devices were programmed, with testing to determine which electrode contacts or contact pairs demonstrated optimal therapeutic efficacy.
For each patient, the contact pair found to provide optimal efficacy was associated with the highest energy in the beta (13-32 Hz) and gamma (48-220 Hz) bands during postoperative LFP recordings at rest and during hand movements. Activities in other frequency bands did not show significant correlations between LFP power and optimal electrode contacts.
Postoperative subband analysis of LFP recordings in beta and gamma frequency ranges may be used to select optimal electrode contacts. These results indicate that LFP recordings from implanted DBS electrodes can provide important clues to guide the optimization of DBS therapy in individual patients.
在美国,治疗帕金森病患者最常用的手术方法是将深部脑刺激(DBS)电极植入丘脑底核内。然而,DBS 设备的程控仍然很困难,并且可能是疗效降低的一个原因。
我们研究了丘脑底核内局部场电位(LFP)活动与程控治疗反应之间的关系。
我们在 4 名帕金森病患者中,在植入后 3 周,在开始长期 DBS 之前,使用单侧放置的微电极记录 LFP。对每个电极的 7 个可能的接触点或接触对进行了计算,每个接触点或接触对在多个 5 到 10 分钟的安静清醒期和 30 秒的手部运动期内进行了计算。随后,对 DBS 设备进行了程控,测试确定哪个电极接触点或接触对显示出最佳治疗效果。
对于每个患者,在术后 LFP 记录中,被发现提供最佳疗效的接触对与休息时和手部运动时的β(13-32 Hz)和γ(48-220 Hz)频段的最高能量相关。其他频段的活动与 LFP 功率之间没有显著的相关性。
术后 LFP 记录的亚频带分析可用于选择最佳电极接触点。这些结果表明,植入 DBS 电极的 LFP 记录可以提供重要线索,指导个体化患者 DBS 治疗的优化。