Efficiency of an enclosed afferent reservoir breathing system during controlled ventilation.

作者信息

Droppert P M, Meakin G, Beatty P C, Mortimer A J, Healy T E

机构信息

University Department of Anaesthesia, Withington Hospital, Manchester.

出版信息

Br J Anaesth. 1991 Jun;66(6):638-42. doi: 10.1093/bja/66.6.638.

Abstract

We describe an enclosed afferent reservoir (EAR) breathing system developed by Ohmeda and designed to operate efficiently in spontaneous and controlled ventilation. The efficiency of the system was evaluated by calculating the fractional utilization of fresh gas in 10 ASA I-III patients during anaesthesia with controlled ventilation. Maximum efficiency occurred when the minute ventilation to fresh gas flow ratio was greater than 1.5. Under these conditions, fractional utilization was relatively constant with a value of 0.73 (95% confidence interval 0.69-0.78). The minimum fresh gas flow for use during controlled ventilation was determined in another eight ASA I-III patients when the minute volume to fresh gas ratio was greater than 1.5. In view of an increased arterial to end-tidal carbon dioxide partial pressure difference in patients in the first part of the study (1.03 kPa), normocapnia was defined as an end-tidal carbon dioxide partial pressure of 4.3 kPa. Normocapnia was achieved with a mean fresh gas flow of 66 ml kg(-1) min(-1), while 70 ml kg(-1) min(-1) produced mild hypocapnia.