Department for Cardiology, University Medical Center Goettingen, Germany.
Europace. 2010 Oct;12(10):1460-6. doi: 10.1093/europace/euq269. Epub 2010 Jul 27.
Ventricular oversensing remains a significant problem in modern implantable cardioverter-defibrillator (ICD) systems. The role of manufacturer-related differences in device-based signal processing on the one hand and sensing lead properties on the other is largely unknown. This retrospective study aimed to evaluate and compare the incidence and mechanisms of ventricular oversensing in single- or dual-chamber ICD systems from Biotronik (BTK), Guidant (GDT), and Medtronic (MDT).
In 245 consecutive patients, device function, stored episodes, and lead parameters were evaluated at implant and during 243 ± 18 days of follow-up. Oversensing occurred in 3.3% of the patients [four T-wave oversensing (TWO), three R-wave double-sensing, and one diaphragmatic myopotentials]. The incidence of ventricular oversensing was generally low if an ICD device and an ICD lead were from one identical manufacturer (BTK: 2/68, GDT: 1/80, and MDT: 1/69; P = 0.67). In the subgroup of ICDs connected to leads from other manufacturers, however, ventricular oversensing was observed in none of 19 GDT and MDT ICDs but in 4 of 9 BTK devices (44.4%, P = 0.012). These four cases of oversensing (three R-wave double-sensing and one TWO) all occurred in BTK ICDs connected to integrated bipolar ICD leads from other manufacturers.
Significant manufacturer-related differences exist in the incidence of ventricular oversensing in single- and dual-chamber ICD systems. Biotronik devices connected to integrated bipolar ICD leads from other manufacturers have an increased risk for R-wave double-sensing.
心室过感知仍然是现代植入式心脏复律除颤器(ICD)系统的一个重大问题。一方面,制造商相关的设备信号处理差异,另一方面是传感器导联特性,其在心室过感知中的作用在很大程度上是未知的。本回顾性研究旨在评估和比较百多力(BTK)、 Guidant(GDT)和美敦力(MDT)的单腔和双腔 ICD 系统中心室过感知的发生率和机制。
在 245 例连续患者中,在植入时和 243±18 天的随访期间评估了设备功能、存储的事件和导联参数。3.3%的患者发生过感知[4 例 T 波过感知(TWO)、3 例 R 波双感知和 1 例膈肌肌电]。如果 ICD 设备和 ICD 导联来自同一制造商(BTK:2/68、GDT:1/80 和 MDT:1/69;P=0.67),则心室过感知的发生率通常较低。然而,在连接到其他制造商的导联的 ICD 亚组中,19 个 GDT 和 MDT ICD 中均未观察到心室过感知,但在 9 个 BTK 设备中有 4 个(44.4%,P=0.012)。这 4 例过感知(3 例 R 波双感知和 1 例 TWO)均发生在与其他制造商的集成双极 ICD 导联相连的 BTK ICD 中。
单腔和双腔 ICD 系统中,心室过感知的发生率存在显著的制造商相关差异。与其他制造商的集成双极 ICD 导联相连的 Biotronik 设备,发生 R 波双感知的风险增加。