Wang S J, Nishimura D G, Macovski A
Magnetic Resonance Systems Research Laboratory, Stanford University, California 94305.
Magn Reson Med. 1991 Jan;17(1):244-51. doi: 10.1002/mrm.1910170127.
We have developed a variation of selective inversion recovery (SIR) angiography that allows us to obtain a collection of several angiograms within the same acquisition time previously required to obtain a single image. In basic SIR, a single readout is performed after the tagging inversion pulse. In multiple-readout SIR, a succession of readout pulses is applied following the inversion pulse. By varying the gradients appropriately during the successive readouts, we can obtain a set of multiple projection-angle angiograms, or, by appropriately spacing the readouts throughout the cardiac cycle, we can obtain a set of time-resolved angiograms. This technique allows us to obtain additional spatial or temporal information without increasing total scan time. A sequence of increasing flip-angle read pulses is used to maintain a constant signal level across the images. A trade-off exists between SNR and the number of images acquired.