Suppr超能文献

气管导管套囊形状对套囊漏液的影响:一项体外研究。

Effect of tracheal tube cuff shape on fluid leakage across the cuff: an in vitro study.

机构信息

Department of Anaesthesia, University Children's Hospital, Steinwiesstrasse 75, CH 8032 Zurich, Switzerland.

出版信息

Br J Anaesth. 2010 Oct;105(4):538-43. doi: 10.1093/bja/aeq202. Epub 2010 Aug 3.

Abstract

BACKGROUND

This study compared the fluid leakage in the new 'tapered' shaped against the classic 'cylindrical' shaped tracheal tube cuffs when placed in different sized tracheas.

METHODS

The 7.5 mm internal diameter (ID) tracheal tube cuffs-Tapered Seal Guard (TSG), Standard Seal Guard (SSG), Hi-Lo, Microcuff, Ruesch, and Portex Profile-were compared in an in vitro apparatus. Vertical artificial tracheas with 16, 20, and 22 mm ID were intubated, 5 ml clear water was applied above the unlubricated tube cuffs, and fluid leakage was measured up to 60 min. Data of tapered vs non-tapered tube cuffs (16 observations) were compared for each tracheal diameter using the Mann-Whitney test.

RESULTS

Median (range) fluid leakage (ml) at 60 min was 2.14 (0.05-4.88), 1.14 (0.00-4.84), and 0.13 (0.00-1.32), respectively, for 16, 20, and 22 mm tracheas in the TSG tube studies when compared with 4.58 (0.44-4.88), 2.21 (0.00-4.81), and 0.00 (0.00-4.81) in the SSG tube and 4.54 (1.54-4.82), 0.90 (0.00-4.49), and 4.85 (4.40-4.99) in the Microcuff tube studies. Leakage in all polyvinylchloride (PVC) tube cuffs was almost complete (5 ml) within 5 min (P<0.001).

CONCLUSIONS

The tapered PU tube cuff was as effective as the cylindrical PU cuffs in smaller tracheal diameters and was more efficient than the cylindrical Microcuff PU tube cuff in larger tracheal diameter in preventing subglottic fluid leakage across the tube cuff tested in this in vitro study. PVC tube cuffs leaked much more and faster than PU cuffs.

摘要

背景

本研究比较了新的“锥形”形状和经典的“圆柱形”形状气管套囊在不同大小的气管中放置时的液体泄漏情况。

方法

在体外装置中比较了 7.5 毫米内径(ID)气管套囊-锥形密封套(TSG)、标准密封套(SSG)、高低型、微型套囊、Ruesch 和 Portex 型-。将内径为 16、20 和 22 毫米的垂直人工气管插管,在未润滑的气管套囊上方应用 5 毫升清水,并测量 60 分钟内的液体泄漏量。使用 Mann-Whitney 检验比较了每个气管直径的锥形与非锥形套囊(16 个观察值)的数据。

结果

在 TSG 管研究中,16、20 和 22 毫米气管的中位(范围)60 分钟液体泄漏(ml)分别为 2.14(0.05-4.88)、1.14(0.00-4.84)和 0.13(0.00-1.32),而 SSG 管分别为 4.58(0.44-4.88)、2.21(0.00-4.81)和 0.00(0.00-4.81),微型套囊管研究中分别为 4.54(1.54-4.82)、0.90(0.00-4.49)和 4.85(4.40-4.99)。所有聚氯乙烯(PVC)管套囊在 5 分钟内几乎完全泄漏(5 毫升)(P<0.001)。

结论

在较小的气管直径中,锥形聚氨基甲酸乙酯套囊与圆柱形聚氨基甲酸乙酯套囊一样有效,在较大的气管直径中,锥形聚氨基甲酸乙酯套囊比圆柱形微型套囊更有效,可防止在体外研究中测试的管套囊下方的声门下液体泄漏。PVC 管套囊泄漏的速度和数量都比 PU 管套囊多。