Suppr超能文献

气管导管套囊周围的液体渗漏:新型微型套囊气管导管的评估

Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube.

作者信息

Dullenkopf Alexander, Gerber Andreas, Weiss Markus

机构信息

Department of Anesthesia, University Children's Hospital, Steinwiesstrasse 75, 8032 Zurich, Switzerland.

出版信息

Intensive Care Med. 2003 Oct;29(10):1849-53. doi: 10.1007/s00134-003-1933-6. Epub 2003 Aug 16.

Abstract

OBJECTIVE

This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-microm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff.

DESIGN

In vitro setup.

MEASUREMENTS AND RESULTS

The following endotracheal tubes (ID 7.5 mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20 mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60 cmH2O. Colored water (5 ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60 min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60 cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10 min only at 10 cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10-30 cmH2O.

CONCLUSIONS

Within the acceptable upper limit for tracheal cuff pressure (25-30 cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings.

摘要

目的

本研究将最近推出的带有超薄(7微米)聚氨酯套囊膜的Microcuff重症监护病房用高容量低压气管内导管与不同制造商生产的气管内导管在套囊周围液体泄漏情况方面进行了比较。

设计

体外实验装置。

测量与结果

比较了以下内径7.5毫米的气管内导管:Mallinckrodt HiLo、Microcuff高容量低压重症监护病房用导管、Portex Profile Soft Seal、Rüsch Super Safety Clear和Sheridan CF。将一个垂直的PVC气管模型(内径20毫米)插管,并将套囊充气至10、15、20、25、30和60厘米水柱。在套囊顶部加入5毫升有色水。记录5、10和60分钟内通过导管套囊泄漏的液体量。每个导管品牌使用两个样本进行了四次实验。在10至60厘米水柱的套囊压力下,所有传统气管内导管均出现套囊周围液体泄漏。在Microcuff导管中,仅在10厘米水柱时10分钟内观察到套囊压力下的液体泄漏。在10 - 30厘米水柱的套囊压力下,Microcuff导管的结果明显优于所有其他导管品牌。

结论

在气管套囊压力的可接受上限(25 - 30厘米水柱)内,Microcuff气管内导管是我们体外实验装置中唯一能防止液体泄漏的受试导管。需要进行体内研究以证实这些发现。