Shokouhi Sepideh, McDonald Benjamin S, Durko Heather L, Fritz Mark A, Furenlid Lars R, Peterson Todd E
Institute of Imaging Science, Vanderbilt University, Nashville, TN 37232-2310 USA.
IEEE Trans Nucl Sci. 2009 Jun 1;56(3):557-564. doi: 10.1109/TNS.2009.2019106.
This work presents characterization studies of thick silicon double-sided strip detectors for a high-resolution small-animal SPECT. The dimension of these detectors is 60.4 mm × 60.4 mm × 1 mm. There are 1024 strips on each side that give the coordinates of the photon interaction, with each strip processed by a separate ASIC channel. Our measurement shows that intrinsic spatial resolution equivalent to the 59 μm strip pitch is attainable. Good trigger uniformity can be achieved by proper setting of a 4-bit DAC in each ASIC channel to remove trigger threshold variations. This is particularly important for triggering at low energies. The thick silicon DSSD (Double-sided strip detector) shows high potential for small-animal SPECT.