通过振荡压力输注实时估计脑脊髓液系统参数。

Real-time estimation of cerebrospinal fluid system parameters via oscillating pressure infusion.

机构信息

Department of Radiation Sciences, Umeå University, 901-85, Umeå, Sweden.

出版信息

Med Biol Eng Comput. 2010 Nov;48(11):1123-31. doi: 10.1007/s11517-010-0670-6. Epub 2010 Aug 6.

Abstract

Hydrocephalus is related to a disturbed cerebrospinal fluid (CSF) system. For diagnosis, lumbar infusion test are performed to estimate outflow conductance, C (out), and pressure volume index, PVI, of the CSF system. Infusion patterns and analysis methods used in current clinical practice are not optimized. Minimizing the investigation time with sufficient accuracy is of major clinical relevance. The aim of this study was to propose and experimentally evaluate a new method, the oscillating pressure infusion (OPI). The non-linear model of the CSF system was transformed into a linear time invariant system. Using an oscillating pressure pattern and linear system identification methods, C (out) and PVI with confidence intervals, were estimated in real-time. Forty-two OPI and constant pressure infusion (CPI) investigations were performed on an experimental CSF system, designed with PVI = 25.5 ml and variable C (out). The ARX model robustly estimated C (out) (mean C (out,OPI) - C (out,CPI) = 0.08 μl/(s kPa), n = 42, P = 0.68). The Box-Jenkins model proved most reliable for PVI (23.7 ± 2.0 ml, n = 42). The OPI method, with its oscillating pressure pattern and new parameter estimation methods, efficiently estimated C (out) and PVI as well as their confidence intervals in real-time. The results from this experimental study show potential for the OPI method and supports further evaluation in a clinical setting.

摘要

脑积水与脑脊液(CSF)系统紊乱有关。为了诊断,进行腰椎输注试验以估计 CSF 系统的流出导纳 C(out)和压力容积指数 PVI。目前临床实践中使用的输注模式和分析方法尚未得到优化。用足够的精度最小化调查时间具有重要的临床意义。本研究的目的是提出并实验评估一种新方法,即振荡压力输注(OPI)。CSF 系统的非线性模型被转化为线性时不变系统。使用振荡压力模式和线性系统识别方法,可以实时估计 C(out)和具有置信区间的 PVI。在设计的 CSF 系统上进行了 42 次 OPI 和恒压输注(CPI)试验,该系统的 PVI = 25.5ml,C(out)可变。ARX 模型可靠地估计了 C(out)(平均 C(out,OPI) - C(out,CPI) = 0.08 μl/(s kPa),n = 42,P = 0.68)。Box-Jenkins 模型最适合 PVI(23.7 ± 2.0 ml,n = 42)。OPI 方法具有振荡压力模式和新的参数估计方法,可实时高效地估计 C(out)和 PVI 及其置信区间。这项实验研究的结果表明 OPI 方法具有潜力,并支持在临床环境中进一步评估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验