Janssen J W, Helbing A R
Klinisch Chemisch Laboratorium St. Franciscus Gasthuis Rotterdam, The Netherlands.
Eur J Clin Chem Clin Biochem. 1991 Mar;29(3):197-201.
The widely used calcium dye reagent, o-cresolphthalein complexone, was replaced by the less well known metallochromic indicator Arsenazo III in a selective multitest analyser. Calibration stability increased from 12 hours to more than 48 hours. With the Arsenazo III reagent we clearly improved the within-lab and between-lab imprecision of the calcium determination. The accuracy of the Arsenazo III method was established with a reference method (atomic absorption spectrophotometry) and good results were obtained with assayed control sera. Furthermore, the mean calcium values of our out-clinic patients remained constant during the change of reagent. The calcium values are influenced by haemolysis; we correct for the observed decrease by adding 0.015 mmol/l Ca per 10 mumol/l Hb. Furthermore they are influenced by very small amounts of Cu ions from the preceding protein determination in the same cuvette. This latter problem is easily overcome by adding 15 mmol/l thiourea to the calcium reagent.
在一台选择性多测试分析仪中,广泛使用的钙染料试剂邻甲酚酞络合酮被鲜为人知的金属铬指示剂偶氮胂III所取代。校准稳定性从12小时提高到了48小时以上。使用偶氮胂III试剂,我们显著改善了实验室内部和实验室之间钙测定的不精密度。通过参考方法(原子吸收分光光度法)确定了偶氮胂III方法的准确性,并且在分析对照血清时获得了良好的结果。此外,在更换试剂期间,我们门诊患者的平均钙值保持恒定。钙值受溶血影响;我们通过每10 μmol/L血红蛋白添加0.015 mmol/L钙来校正观察到的降低。此外,它们还受到同一比色皿中先前蛋白质测定产生的极少量铜离子的影响。通过向钙试剂中添加15 mmol/L硫脲,很容易克服后一个问题。