用于血清和血浆中钙的光学稳健测量的单一稳定试剂(偶氮胂III)。

Single stable reagent (Arsenazo III) for optically robust measurement of calcium in serum and plasma.

作者信息

Leary N O, Pembroke A, Duggan P F

机构信息

Biochemistry Laboratory, Regional Hospital, Cork, Ireland.

出版信息

Clin Chem. 1992 Jun;38(6):904-8.

PMID:1597016
Abstract

We describe a method based on a single stable reagent for the determination of calcium in serum and plasma with use of Arsenazo III, 200 mumol/L in 50 mmol/L 1,4-piperazinediethanesulfonic acid (PIPES) buffer. The method showed significant positive interference in plasma at pH less than 6.6 because of the precipitation of fibrinogen, which was eliminated by increasing the pH to 6.8. The assay showed no interference from as much as 600 mumol of bilirubin and 12 g of hemoglobin per liter when applied in a simple monochromatic procedure at 660 nm. The standard curve for calcium was linear from 0 to 5.0 mmol/L. Addition of Intralipid at concentrations greater than 3 g/L demonstrated positive interference, which could be eliminated by using a 700-nm blanking wavelength. The procedure showed good agreement with all-method mean values from two external quality-control schemes.

摘要

我们描述了一种基于单一稳定试剂的方法,该方法使用偶氮胂III在50 mmol/L 1,4 - 哌嗪二乙磺酸(PIPES)缓冲液中(浓度为200 μmol/L)测定血清和血浆中的钙。该方法在pH小于6.6时,由于纤维蛋白原沉淀,在血浆中显示出显著的正干扰,通过将pH提高到6.8可消除这种干扰。当在660 nm处采用简单的单色程序时,该测定法不受每升高达600 μmol胆红素和12 g血红蛋白的干扰。钙的标准曲线在0至5.0 mmol/L范围内呈线性。添加浓度大于3 g/L的脂肪乳剂显示出正干扰,通过使用700 nm的空白波长可以消除这种干扰。该程序与两个外部质量控制方案的所有方法平均值显示出良好的一致性。