评估大腿肌肉横截面积和衰减时 CT 测量的可靠性。

Reliability of computed tomography measurements in assessment of thigh muscle cross-sectional area and attenuation.

机构信息

Department of Radiology, Karolinska University Hospital, Karolinska Institutet, Stockholm, Sweden.

出版信息

BMC Med Imaging. 2010 Aug 11;10:18. doi: 10.1186/1471-2342-10-18.

Abstract

BACKGROUND

Advancement in technology of computer tomography (CT) and introduction of new medical imaging softwares enables easy and rapid assessment of muscle cross-sectional area (CSA) and attenuation. Before using these techniques in clinical studies there is a need for evaluation of the reliability of the measurements. The purpose of the study was to evaluate the inter- and intra-observer reliability of ImageJ in measuring thigh muscles CSA and attenuation in patients with anterior cruciate ligament (ACL) injury by computer tomography.

METHODS

31 patients from an ongoing study of rehabilitation and muscle atrophy after ACL reconstruction were included in the study. Axial CT images with slice thickness of 10 mm at the level of 150 mm above the knee joint were analyzed by two investigators independently at two times with a minimum of 3 weeks between the two readings using NIH ImageJ. CSA and the mean attenuation of individual thigh muscles were analyzed for both legs.

RESULTS

Mean CSA and mean attenuation values were in good agreement both when comparing the two observers and the two replicates. The inter- and intraclass correlation (ICC) was generally very high with values from 0.98 to 1.00 for all comparisons except for the area of semimembranosus. All the ICC values were significant (p < 0,001). Pearson correlation coefficients were also generally very high with values from 0.98 to 1.00 for all comparisons except for the area of semimembranosus (0.95 for intraobserver and 0.92 for interobserver).

CONCLUSION

This study has presented ImageJ as a method to monitor and evaluate CSA and attenuation of different muscles in the thigh using CT-imaging. The method shows an overall excellent reliability with respect to both observer and replicate.

摘要

背景

计算机断层扫描(CT)技术的进步和新的医学成像软件的引入使得肌肉横截面积(CSA)和衰减的快速评估变得更加容易。在将这些技术应用于临床研究之前,需要评估测量的可靠性。本研究的目的是评估 ImageJ 在通过 CT 测量前交叉韧带(ACL)损伤患者大腿肌肉 CSA 和衰减时的观察者内和观察者间可靠性。

方法

31 名来自 ACL 重建后康复和肌肉萎缩的正在进行的研究的患者被纳入本研究。使用 NIH ImageJ,两名研究人员在两次测量时分别独立分析膝关节上方 150mm 处厚度为 10mm 的轴向 CT 图像,两次测量之间至少间隔 3 周。分析了每条腿的 CSA 和各个大腿肌肉的平均衰减值。

结果

当比较两名观察者和两次重复测量时,CSA 和平均衰减值之间存在良好的一致性。除半膜肌外,所有比较的观察者内和观察者间的组内相关系数(ICC)均非常高,为 0.98 至 1.00。所有 ICC 值均具有统计学意义(p < 0.001)。除半膜肌外,Pearson 相关系数也普遍非常高,为 0.98 至 1.00(观察者内为 0.95,观察者间为 0.92)。

结论

本研究提出了使用 CT 成像监测和评估大腿不同肌肉 CSA 和衰减的 ImageJ 方法。该方法在观察者和重复测量方面均表现出总体上极好的可靠性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/3c2e/2928755/d24d11c3eac8/1471-2342-10-18-1.jpg