Suppr超能文献

基于大邻域加权强度平均的低剂量腹部 CT 图像质量改善。

Improving low-dose abdominal CT images by Weighted Intensity Averaging over Large-scale Neighborhoods.

机构信息

The Laboratory of Image Science and Technology, Southeast University, China.

出版信息

Eur J Radiol. 2011 Nov;80(2):e42-9. doi: 10.1016/j.ejrad.2010.07.003. Epub 2010 Aug 14.

Abstract

PURPOSE

Though highly desirable in radiologic procedures, low-dose CT (LDCT) images tend to be severely degraded by quantum noise and non-stationary artifacts. The purpose of this paper is to improve the abdominal LDCT images by the approach of Weighted Intensity Averaging over Large-scale Neighborhoods (WIA-LN).

MATERIALS AND METHODS

In the implementation of the proposed WIA-LN method, the processed pixel intensities are adaptively calculated as the weighted intensity averaging of the pixels with similar surrounding structures throughout a large-scale neighborhood. Both phantom and clinical abdominal CT images from a 16 detector rows Siemens CT were acquired at standard and 80% reduced tube current time products (150 mAs and 30 mAs corresponding to standard-dose and low-dose protocols, respectively). Visual comparison, statistical qualitative analysis (image quality scores and hepatic cyst diagnosis), and quantitative calculation (noise and contrast-to-noise ratio) are made.

RESULTS

Better vision and quantitative performance are realized using the proposed WIA-LN method. Compared to original LDCT and standard-dose CT (SDCT) images, statistically significant improvement of noise/artifacts suppression, contrast preservation and hepatic cyst detection in LDCT images are achieved by using the proposed method (P<0.05).

CONCLUSION

With the tube current reduced to approximate one-fifth of the standard tube current setting, clinically acceptable images can still be obtained by using the proposed method.

摘要

目的

虽然在放射学程序中非常理想,但低剂量 CT(LDCT)图像往往会因量子噪声和非平稳伪影而严重降级。本文旨在通过大区域邻域加权强度平均(WIA-LN)方法来改善腹部 LDCT 图像。

材料与方法

在提出的 WIA-LN 方法的实施中,处理后的像素强度作为具有相似周围结构的像素的加权强度平均自适应地计算。在标准和 80%降低管电流时间乘积(分别对应标准剂量和低剂量方案的 150 mAs 和 30 mAs)下,使用来自 16 排西门子 CT 的体模和临床腹部 CT 图像。进行视觉比较、统计定性分析(图像质量评分和肝囊肿诊断)和定量计算(噪声和对比噪声比)。

结果

使用提出的 WIA-LN 方法实现了更好的视觉和定量性能。与原始 LDCT 和标准剂量 CT(SDCT)图像相比,使用该方法在 LDCT 图像中实现了噪声/伪影抑制、对比度保持和肝囊肿检测的统计学显著改善(P<0.05)。

结论

通过将管电流降低到标准管电流设置的约五分之一,使用该方法仍可获得临床可接受的图像。