Allison D P, Thompson J R, Jacobson K B, Warmack R J, Ferrell T L
Oak Ridge National Laboratory, TN 37831.
Scanning Microsc. 1990 Sep;4(3):517-22.
We present scanning tunneling microscope (STM) images of uncoated deoxyribonucleic acid (DNA) electrochemically mounted on highly-oriented pyrolytic graphite (HOPG) and imaged in air. Images of linear abnormalities inherent to HOPG surfaces that can be confused with DNA are also presented. Scanning tunneling spectroscopic (STS) images generated by superimposing a small, high frequency ac bias onto the dc tunnel bias and recording the ac current signal were taken simultaneously with the topographic images. These spectroscopic images reveal contrast due to local conductivity variations and can be used to differentiate DNA molecules from graphite artifacts.
我们展示了电化学固定在高度定向热解石墨(HOPG)上并在空气中成像的未包覆脱氧核糖核酸(DNA)的扫描隧道显微镜(STM)图像。还展示了HOPG表面固有的可能与DNA混淆的线性异常图像。通过在直流隧道偏压上叠加一个小的高频交流偏压并记录交流电流信号生成的扫描隧道光谱(STS)图像与形貌图像同时采集。这些光谱图像揭示了由于局部电导率变化而产生的对比度,可用于区分DNA分子和石墨伪像。