Int J Audiol. 2010 Dec;49(12):868-76. doi: 10.3109/14992027.2010.503664. Epub 2010 Sep 3.
Audiometric measurements through a hearing aid ('in-situ') may facilitate provision of hearing services where these are limited. This study investigated the validity and reliability of in-situ air conduction hearing thresholds measured with closed and open domes relative to thresholds measured with insert earphones, and explored sources of variability in the measures. Twenty-four adults with sensorineural hearing impairment attended two sessions in which thresholds and real-ear-to-dial-difference (REDD) values were measured. Without correction, significantly higher low-frequency thresholds in dB HL were measured in-situ than with insert earphones. Differences were due predominantly to differences in ear canal SPL, as measured with the REDD, which were attributed to leaking low-frequency energy. Test-retest data yielded higher variability with the closed dome coupling due to inconsistent seals achieved with this tip. For all three conditions, inter-participant variability in the REDD values was greater than intra-participant variability. Overall, in-situ audiometry is as valid and reliable as conventional audiometry provided appropriate REDD corrections are made and ambient sound in the test environment is controlled.
通过助听器(“原位”)进行测听可能有助于在听力服务有限的情况下提供听力服务。本研究调查了与插入式耳机测量相比,使用封闭和开放耳塞测量的原位空气传导听力阈值的有效性和可靠性,并探讨了测量值中可变性的来源。24 名患有感音神经性听力障碍的成年人参加了两次测试,其中测量了阈值和真耳到耳罩差(REDD)值。未经校正,原位测量的低频阈值以分贝毫瓦(dB HL)表示,显著高于插入式耳机。差异主要归因于使用 REDD 测量的耳道声压级(SPL)的差异,这归因于低频能量泄漏。由于使用这种探头无法实现一致的密封,因此测试-重测数据显示出更高的封闭圆顶耦合的可变性。对于所有三种情况,REDD 值的组间变异性大于个体内变异性。总体而言,只要进行适当的 REDD 校正并控制测试环境中的环境噪声,原位测听与传统测听一样有效且可靠。