Funama Yoshinori, Awai Kazuo, Hatemura Masahiro, Shimamura Masamitchi, Yanaga Yumi, Oda Seitaro, Yamashita Yasuyuki
Department of Radiological Sciences, School of Health Sciences, Kumamoto University, Kuhonji, Kumamoto, Japan.
Radiol Phys Technol. 2008 Jan;1(1):33-7. doi: 10.1007/s12194-007-0005-y. Epub 2007 Nov 1.
To investigate whether it is possible to obtain adequate images at uniform image noise levels and reduced radiation exposure with our automatic tube current modulation (ATCM) technique for 64-detector CT. The study population consisted of 64 patients with known or suspected lung or abdominal disease. We used a 64-detector CT scanner (LightSpeed VCT, GE Healthcare, Waukesha, WI, USA) and a combined angular and longitudinal tube current modulation technique (Smart mA, GE Healthcare, Waukesha, WI, USA) to examine 34 patients. The scanning parameters were identical; the minimum and maximum tube current thresholds were 50 and 800 mA, respectively. For study of the constant tube current technique, 30 additional patients were examined at 350 mA. The CT number and image noise (SD of the CT number) were measured in the 64 patients at six levels, i.e., the center of the left ventricle, the liver dome, the porta hepatis, the center of the spleen and the right and left renal pelvis. When we used the ATCM technique, the mean image noise ranged from 8.40 at the center of the left ventricle to 11.31 at the porta hepatis; the mean tube current ranged from 105.9 mAs at the center of the left ventricle to 169.6 mAs at the center of the spleen. The mean dose reduction rate per constant tube current at 175 mAs ranged from 3.1 to 39.5%. By use of the ATCM technique, it is possible to maintain a constant image noise level with a 64-detector CT.
为了研究使用我们的自动管电流调制(ATCM)技术在64层CT上是否能够在均匀的图像噪声水平下获得足够的图像并减少辐射暴露。研究人群包括64例已知或疑似患有肺部或腹部疾病的患者。我们使用64层CT扫描仪(LightSpeed VCT,GE Healthcare,美国威斯康星州沃基沙)和一种角度与纵向相结合的管电流调制技术(Smart mA,GE Healthcare,美国威斯康星州沃基沙)对34例患者进行检查。扫描参数相同;管电流的最小和最大阈值分别为50和800 mA。对于恒定管电流技术的研究,另外30例患者以350 mA进行检查。在64例患者的六个层面,即左心室中心、肝顶、肝门、脾中心以及左右肾盂处测量CT值和图像噪声(CT值的标准差)。当我们使用ATCM技术时,平均图像噪声范围从左心室中心的8.40到肝门处的11.31;平均管电流范围从左心室中心的105.9 mAs到脾中心的169.6 mAs。在175 mAs时,相对于恒定管电流的平均剂量降低率范围为3.1%至39.5%。通过使用ATCM技术,使用64层CT有可能维持恒定的图像噪声水平。