Zheng Pei-chao, Wang Hong-mei, Li Jian-quan, Han Hai-yan, Xi Xiao-qin, Chu Yan-nan
Laboratory of Environmental Spectroscopy, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China.
Guang Pu Xue Yu Guang Pu Fen Xi. 2010 Jul;30(7):1948-51.
Toxic metal elements in waters and wastewaters contaminate the environment and greatly threaten the health of human beings, therefore developing a rapid monitor for metal residues in aqueous solutions is urgently required. In the present work, a new homemade apparatus of electrolyte cathode atmospheric glow discharge emission spectroscopy was developed and described. It can detect and discriminate many kinds of trace mental elements by atomic emission spectrum from atmospheric pressure liquid cathode glow discharge. In order to estimate the analytical performance of the present atmospheric pressure electrolyte cathode glow discharge emission spectroscopy system, the detection limit values for Na, Li, Cu, Pb and Mn were obtained based on 3sigma of the background signal, and the current limits of detection were 0.008, 0.005, 1.1, 2.06 and 1.95 mg L(-1), respectively. It demonstrates that the atmospheric pressure electrolyte cathode glow discharge emission spectroscopy has a promising application in real time measurements of metal residues in aqueous solutions.
水体和废水中的有毒金属元素污染环境,对人类健康构成重大威胁,因此迫切需要开发一种能快速监测水溶液中金属残留量的方法。在本研究中,开发并描述了一种新型的自制电解质阴极常压辉光放电发射光谱仪。它可以通过常压液体阴极辉光放电产生的原子发射光谱检测和鉴别多种痕量金属元素。为评估当前常压电解质阴极辉光放电发射光谱系统的分析性能,基于背景信号的3σ得出了Na、Li、Cu、Pb和Mn的检出限,其电流检出限分别为0.008、0.005、1.1、2.06和1.95 mg L(-1)。结果表明,常压电解质阴极辉光放电发射光谱在水溶液中金属残留量的实时测定方面具有广阔的应用前景。