American Red Cross Holland Laboratory, Rockville, Maryland 20855, USA.
Transfusion. 2011 Mar;51(3):636-42. doi: 10.1111/j.1537-2995.2010.02869.x. Epub 2010 Sep 16.
Apheresis platelet (PLT) units are not routinely agitated during transit. Our study compared the in vitro properties of apheresis PLT units that were stored with continuous agitation (CA) and without continuous agitation (WCA) during two separate periods, immediately after collection and between Day 2 and Day 3 of storage.
Two identical apheresis PLTs units were prepared from collections with Amicus (n = 11, Fenwal, Inc.) and Trima (n = 10, CaridianBCT) cell separators. One apheresis PLT unit was continuously agitated, starting routinely within 30 minutes of collection, and an identical apheresis PLT unit was held without agitation initially for 7 to 8 hours and subsequently for 24 hours between Day 2 and Day 3 of storage. The apheresis PLT units were maintained WCA at 20 to 24 °C in a shipping box. In vitro PLT properties were evaluated on Day 1 (day after collection), after 5 and 7 days of storage.
With both Amicus and Trima apheresis PLT units, the mean PLT content and concentration of CA and WCA were comparable and essentially constant throughout storage. Mean pH levels (± 1 SD) after 5 days for Amicus apheresis PLT units were 6.97 ± 0.20 (WCA) and 7.13 ± 0.16 (p < 0.001, CA) and for Trima apheresis PLT units 6.97 ± 0.21 (WCA) and 7.22 ± 0.17 (p < 0.001, CA). In vitro variables, including percentage of disc PLTs, extent of shape change, and hypotonic stress levels, after 5 days of storage, showed mean differences between WCA and CA that were less than 15%.
The in vitro results show that apheresis PLT units can be stored without agitation for 7 to 8 hours immediately after collection and also subsequently during storage for 24 hours with minimal influence on in vitro PLT properties compared to continuously agitated PLTs.
在运输过程中,并非常规对机采血小板(PLT)单位进行搅拌。我们的研究比较了在两个不同时间段内储存的机采 PLT 单位的体外特性,一个在采集后立即进行连续搅拌(CA),另一个在采集后立即不进行连续搅拌(WCA),然后在储存的第 2 天至第 3 天之间再进行 24 小时的 WCA。
用 Amicus(n = 11,Fenwal,Inc.)和 Trima(n = 10,CaridianBCT)细胞分离机采集了 2 个相同的机采 PLT 单位。一个机采 PLT 单位从采集开始就进行连续搅拌,通常在采集后 30 分钟内开始,而另一个机采 PLT 单位最初不进行搅拌,持续 7 至 8 小时,然后在储存的第 2 天至第 3 天之间再进行 24 小时的 WCA。机采 PLT 单位在 20 至 24°C 的运输箱中进行 WCA。在第 1 天(采集后第 1 天)、第 5 天和第 7 天评估机采 PLT 的体外特性。
使用 Amicus 和 Trima 机采 PLT 单位,CA 和 WCA 的 PLT 含量和浓度的平均值在整个储存过程中是相似的,基本保持不变。Amicus 机采 PLT 单位在第 5 天后的平均 pH 值(±1 SD)为 6.97 ± 0.20(WCA)和 7.13 ± 0.16(p < 0.001,CA),Trima 机采 PLT 单位为 6.97 ± 0.21(WCA)和 7.22 ± 0.17(p < 0.001,CA)。在储存 5 天后,体外变量,包括盘状 PLT 的百分比、形状变化程度和低渗应激水平,WCA 和 CA 之间的平均差异小于 15%。
体外结果表明,机采 PLT 单位在采集后立即不进行搅拌,可保持 7 至 8 小时,随后在储存期间再进行 24 小时的 WCA,与连续搅拌的 PLT 相比,对体外 PLT 特性的影响最小。