School of Health, Glasgow Caledonian University, Glasgow, UK.
Spinal Cord. 2011 Mar;49(3):445-50. doi: 10.1038/sc.2010.126. Epub 2010 Sep 21.
Keeping physically active is important for people who mobilize using a wheelchair. However, current tools to measure physical activity in the wheelchair are either not validated or limited in their application. The purpose of this study was to develop and validate a monitoring system to measure wheelchair movement.
The system developed consisted of a tri-axial accelerometer placed on the wheel of a wheelchair and an analysis algorithm to interpret the acceleration signals. The two accelerometer outputs in the plane of the wheel were used to calculate the angle of the wheel. From this, outcome measures of wheel revolutions, absolute angle and duration of movement were derived and the direction of movement (forwards or backwards) could be distinguished. Concurrent validity was assessed in comparison with video analysis in 14 people with spinal cord injury using their wheelchair on an indoor track and outdoor wheelchair skills course. Validity was assessed using intraclass correlation coefficients (ICC(2,1)) and Bland-Altman plots.
The monitoring system demonstrated excellent validity for wheel revolutions, absolute angle and duration of movement (ICC(2,1)>0.999, 0.999, 0.981, respectively) in both manual and powered wheelchairs, when the wheelchair was propelled forwards and backwards, and for movements of various durations.
This study has found this monitoring system to be an accurate and objective tool for measuring detailed information on wheelchair movement and maneuvering regardless of the propulsion technique, direction and speed.
对于使用轮椅移动的人来说,保持身体活跃很重要。然而,目前用于测量轮椅活动的工具要么未经验证,要么应用范围有限。本研究的目的是开发和验证一种监测系统,以测量轮椅运动。
开发的系统包括一个三轴加速度计,放置在轮椅的轮子上,以及一个用于解释加速度信号的分析算法。轮子平面上的两个加速度计输出用于计算轮子的角度。由此,得出轮子旋转次数、绝对角度和运动持续时间的结果测量值,并可以区分运动方向(前进或后退)。在室内轨道和户外轮椅技能课程上,使用轮椅对 14 名脊髓损伤患者进行了视频分析的同时,评估了该监测系统的同时效度。使用组内相关系数(ICC(2,1))和 Bland-Altman 图评估了效度。
无论推进技术、方向和速度如何,该监测系统在手动和动力轮椅上,以及在向前和向后推动轮椅时,对于各种持续时间的运动,都表现出了对轮子旋转次数、绝对角度和运动持续时间的优异有效性(ICC(2,1)>0.999、0.999、0.981,分别)。
本研究发现,该监测系统是一种准确、客观的工具,可用于测量轮椅运动和操纵的详细信息,无论推进技术、方向和速度如何。