通过应变计体积描记法测量静脉泵功能。

Measuring venous pumping function by strain-gauge plethysmography.

作者信息

Mosti G, Partsch H

机构信息

Angiology Department, Barbantini Hospital, Lucca, Italy.

出版信息

Int Angiol. 2010 Oct;29(5):421-5.

Abstract

AIM

Hemodynamic effects of compression devices in patients with venous insufficiency are still under dispute. The aim of this paper was to describe a method which is able to give quantitative results concerning an improvement of the venous pumping function by using compression devices in patients with venous insufficiency.

METHODS

The ejection fraction (EF) of the calf pump can be measured using strain gauge plethysmography proximal to a compression device on the leg and the degree of improvement depending on the exerted pressure and the elastic property of the material can be assessed. Experiments were carried out in 15 normal volunteers and in a total of 68 patients with severe superficial venous incompetence and who were investigated without compression and with various compression devices measuring interface pressure in different body positions.

RESULTS

Examples demonstrating the usefulness of the plethysmographic method are given. The median values of EF were 64.6 (IQR 63.4-68.6) in normal subjects and about half in the venous patients. Statistically significant improvement was achieved using inelastic bandages, resulting in further improvements with increasing pressures between 20 and 60 mmHg. Elastic stockings led only to a moderate improvement of the deficient venous pumping function. Significant correlations were found between EF and standing and walking pressure.

CONCLUSION

The described plethysmographic method is very useful to assess haemodynamic effects of different compression devices in patients with venous insufficiency.

摘要

目的

加压装置对静脉功能不全患者的血流动力学影响仍存在争议。本文的目的是描述一种方法,该方法能够通过对静脉功能不全患者使用加压装置来给出关于静脉泵功能改善的定量结果。

方法

可以使用腿部加压装置近端的应变片体积描记法测量小腿泵的射血分数(EF),并可以评估取决于施加压力和材料弹性特性的改善程度。对15名正常志愿者以及总共68名患有严重浅静脉功能不全的患者进行了实验,这些患者在未使用加压装置以及使用各种加压装置的情况下接受了不同体位下的界面压力测量。

结果

给出了证明体积描记法有用性的实例。正常受试者的EF中位数为64.6(四分位距63.4 - 68.6),静脉疾病患者的EF中位数约为正常受试者的一半。使用无弹性绷带可实现统计学上的显著改善,随着压力在20至60 mmHg之间增加,改善进一步提高。弹性长袜仅使不足的静脉泵功能有适度改善。发现EF与站立和行走压力之间存在显著相关性。

结论

所描述的体积描记法对于评估不同加压装置对静脉功能不全患者的血流动力学影响非常有用。