Department of Clinical Biochemistry, University College London Hospitals, 60 Whitfield Street, London, UK.
Ann Clin Biochem. 2010 Nov;47(Pt 6):564-6. doi: 10.1258/acb.2010.010068. Epub 2010 Oct 6.
In an evaluation of androstenedione results from patient serum samples using the Siemens Immulite 2500 analyser and manual Coat-A-Count (CAC) methods, three outliers were evident with grossly elevated results in the CAC assay.
The clinic notes of three patients with apparently high serum androstenedione concentrations by the CAC assay were checked for medications. The samples were all from patients with polycystic ovary syndrome taking 100-200 mg/d of a steroidal antiandrogen (spironolactone). Two other patients on 50 mg spironolactone per day had less markedly higher androstendione results with the CAC assay. In a further five patients who were selected since they were on spironolactone and had high androstenedione results by the CAC method, spironolactone was temporarily withdrawn and fresh blood samples obtained for analysis.
Spironolactone treatment was associated with higher androstenedione concentrations measured by the CAC assay that reverted to normal on treatment withdrawal. Based on a single test with spironolactone at 1000 ng/mL, the manufacturer reported only 0.109% interference in the CAC assay.
Spironolactone (and/or its metabolites) may interfere in the Siemens CAC assay for androstenedione but not in the Immulite 2500 assay. This experience highlights the need for information from clinicians on drug treatment when laboratory investigations are requested. Drug interferences in immunoassay are common and need evaluation beyond tests performed to certify laboratory reagents.
在使用西门子 Immulite 2500 分析仪和手动 Coat-A-Count(CAC)方法对患者血清样本中的雄烯二酮结果进行评估时,CAC 测定法中明显存在三个异常值,结果显著升高。
检查了 CAC 测定法中血清雄烯二酮浓度明显升高的三个患者的临床记录,以检查药物情况。所有样本均来自服用 100-200mg/d 甾体抗雄药物(螺内酯)的多囊卵巢综合征患者。另外两名每天服用 50mg 螺内酯的患者 CAC 测定法的雄烯二酮结果略高。另外,还选择了另外五名因服用螺内酯且 CAC 方法的雄烯二酮结果较高而停药的患者,暂时停止服用螺内酯并采集新鲜血液样本进行分析。
螺内酯治疗与 CAC 测定法测量的雄烯二酮浓度升高有关,停药后恢复正常。根据在 1000ng/mL 螺内酯的单次测试,制造商报告 CAC 测定法的干扰仅为 0.109%。
螺内酯(和/或其代谢物)可能会干扰西门子 CAC 测定法中的雄烯二酮,但不会干扰 Immulite 2500 测定法。这种经验强调了在要求进行实验室检查时,需要从临床医生那里获取有关药物治疗的信息。免疫测定中的药物干扰很常见,需要在实验室试剂认证测试之外进行评估。